Physio-Control LIFEPAK 15

Error: TH_ERROR_PA_RATE_OUT_OF_RANGE Pacing Rate Out Of Tolerance

On this page

Asset Type

Defibrillator

Manufacturer

Physio-Control

Model

LIFEPAK 15

What This Guide Helps With

Troubleshooting TH_ERROR_PA_RATE_OUT_OF_RANGE when LIFEPAK 15 pacing rate output does not match the selected pacing rate.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Remove the Device From Service

Check External and Simple Causes First

Verify Power Condition

Confirm the Issue Is Not Accessory-Related

Do Not Attempt Clinical Use Testing

Process the Therapy Service Code

Perform Pacer Characteristic Testing After Corrective Action

Return to Service Only After Passing Verification

If the Problem Persists

If TH_ERROR_PA_RATE_OUT_OF_RANGE or therapy service code a010 remains present after basic external checks, known-good accessories, and controlled bench evaluation, the likely condition is an internal pacer/therapy control fault.

Remove the device from service, label it Out of Service, and send it for repair or formal bench evaluation.

This is not a normal user-correctable issue. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot pacing output problems on an active patient. Transfer pacing/defibrillation readiness to another verified defibrillator before evaluating the LIFEPAK 15 so therapy continuity is maintained.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"LIFEPAK 15 reported TH_ERROR_PA_RATE_OUT_OF_RANGE with therapy service code a010. Pacing rate was reported out of tolerance, with selected pacing rate not matching actual rate."

Cause

What was observed during troubleshooting.

Example:
"External inspection and basic power/accessory checks did not identify a correctable external issue. Error indicates a pacer rate tolerance fault requiring corrective action and pacer characteristic verification."

Resolution

What action was taken.

Example:
"Removed LIFEPAK 15 from service and labeled Out of Service. Documented therapy service code a010 and sent unit for repair/bench evaluation. Pacer characteristic testing required before return to use."

Helpful Details to Include (If Known)

Final Thought

A pacing rate out-of-tolerance error on a LIFEPAK 15 should be treated as a therapy function reliability issue. Clinical Engineering should confirm the reported error, rule out obvious external causes, remove the device from service, and require pacer characteristic testing after corrective action before the device returns to patient care.

That is successful troubleshooting.