Physio-Control LIFEPAK 15

Error: TH_ERROR_SHOCK_NOT_DELIVERED Shock Not Delivered

On this page

Asset Type

Defibrillator

Manufacturer

Physio-Control

Model

LIFEPAK 15

What This Guide Helps With

This guide helps troubleshoot TH_ERROR_SHOCK_NOT_DELIVERED when the Physio-Control LIFEPAK 15 reports that a shock was not delivered or therapy output may not have occurred. Because this error directly involves defibrillation therapy, it should be treated as a high-priority patient safety issue until the device passes controlled delivered-energy verification.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Remove the Device From Service

Check External Therapy Setup During Controlled Testing Only

Inspect the Therapy Cable

Verify Electrodes or Test Load Setup

Check for Accessory-Related Failure

Check Power and Battery Condition

Review Related Service Indicators or Therapy Codes

Perform Controlled Defibrillator Output Verification After Corrective Action

Stop When Resolved

If the Problem Persists

If TH_ERROR_SHOCK_NOT_DELIVERED remains after external therapy cable, electrode, test load, power, and setup issues are ruled out, treat the condition as an internal therapy delivery failure.

The device should remain removed from service, labeled Out of Service, escalated for repair or manufacturer-level service evaluation, and held from use until corrective action and delivered-energy testing are completed.

Persistent shock-not-delivered errors strongly indicate a device-level therapy subsystem fault. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot this error on an active patient. Move the patient to a backup defibrillator first, maintain therapy continuity, and do not rely on a LIFEPAK 15 with an unresolved shock delivery error for emergency therapy, cardioversion, or code response.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"LIFEPAK 15 reported TH_ERROR_SHOCK_NOT_DELIVERED with therapy service code a00b. Staff reported shock was not delivered during therapy attempt or operational checkout."

Cause

What was observed during troubleshooting.

Example:
"External therapy setup was inspected. Therapy cable, electrodes or test load, power source, and controlled test setup were checked. Error indicates shock delivery failure; persistent condition suggests internal therapy subsystem fault if not resolved by accessory correction."

Resolution

What action was taken.

Example:
"Removed LIFEPAK 15 from service and labeled Out of Service. Verified therapy cable, electrode, and test load setup during controlled bench testing. Device escalated for corrective action and defibrillator delivered-energy testing prior to return to service."

Helpful Details to Include (If Known)

Final Thought

A shock-not-delivered error on a defibrillator is a high-priority patient safety issue. Clinical Engineering should confirm the reported condition, rule out obvious external therapy cable or test setup problems, and avoid assuming the unit is safe just because the message clears. If the error persists, escalation and delivered-energy verification are required before the device can return to patient use.

That is successful troubleshooting.