Physio-Control LIFEPAK 15

Error: PA_ERROR_ROM_TEST ROM CRC Test Failed

On this page

Asset Type

Defibrillator

Manufacturer

Physio-Control

Model

LIFEPAK 15

What This Guide Helps With

This guide helps troubleshoot PA_ERROR_ROM_TEST with therapy service code a213 on the Physio-Control LIFEPAK 15. This error indicates a ROM CRC test failure and may affect therapy system reliability.

A ROM CRC failure should be treated as a serious therapy system integrity warning. The device should not be used for patient monitoring, pacing, cardioversion, or defibrillation until the condition is evaluated and the unit passes required checks.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Remove the Device From Service

Check Basic External Conditions First

Recheck With a Known-Good Battery

Inspect External Condition of the Device

Process and Review the Service Log

Perform a Controlled Power Cycle Only After Safety Isolation

Run Required Operator or Performance Checks

Determine Whether the Fault Is Device-Related

If the Problem Persists

If PA_ERROR_ROM_TEST / service code a213 persists after basic external checks, known-good battery testing, controlled restart, and service log review, external and common causes have been ruled out.

A persistent ROM CRC test failure strongly indicates an internal therapy system or stored-code integrity fault.

The LIFEPAK 15 should be removed from service, labeled Out of Service, sent for authorized repair or bench evaluation, and documented with the service code and service log findings.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot this error on an active patient. Move patient care to another verified defibrillator before evaluating the LIFEPAK 15. A defibrillator with an unresolved therapy service code should not be relied on for emergency therapy, pacing, cardioversion, or monitoring.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"User reported LIFEPAK 15 displayed PA_ERROR_ROM_TEST with therapy service code a213. Service information indicated ROM CRC test failed. Device removed from patient use for evaluation."

Cause

What was observed during troubleshooting.

Example:
"Known-good battery and external power checks did not identify an external power or accessory cause. Service log confirmed PA_ERROR_ROM_TEST / a213 condition, indicating likely internal therapy system or ROM integrity fault."

Resolution

What action was taken.

Example:
"Processed service log, verified fault condition, removed LIFEPAK 15 from service, labeled Out of Service, and escalated unit for repair/bench evaluation. Device not returned to clinical use."

Helpful Details to Include (If Known)

Final Thought

PA_ERROR_ROM_TEST on the LIFEPAK 15 should be treated as a serious therapy system integrity warning. Clinical Engineering should confirm the reported code, rule out simple external power issues, review the service log, and avoid unnecessary internal disassembly. If the error persists, escalation is the correct and safest action. Clear CCR documentation protects patient safety, supports repair decisions, and shows that troubleshooting was logical and complete.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide