Physio-Control LIFEPAK 20 / 20e

ECG Display Errors or Artifacts

On this page

Asset Type

Defibrillator

Manufacturer

Physio-Control

Model

LIFEPAK 20 / 20e

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting ECG display errors or artifacts on a LIFEPAK 20 / 20e defibrillator. Symptoms may include unusual waveforms, baseline wander, noisy signal, intermittent ECG display, or distorted readings. The focus is on external and easily verifiable factors before considering internal failure.

Step-by-Step Troubleshooting

Check Patient Connection

Why: Poor contact or loose electrodes are the most common cause of artifacts.

Inspect ECG Cables

Why: Damaged or loose cables can introduce intermittent or noisy signals.

Check Electrode Quality and Expiration

Why: Electrodes lose conductivity over time, causing artifacts.

Verify Device Settings

Why: Incorrect configuration can produce abnormal display or baseline wander.

Assess Environmental Interference

Why: EMI or motion artifacts can distort the ECG waveform.

Perform Device Self-Test

Why: Confirms whether the device circuitry and display are functioning correctly.

If the Problem Persists

Common external causes have been ruled out; the ECG module or internal circuitry may be faulty.

Reminder: Knowing when to escalate is part of proper troubleshooting.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“ECG waveform is showing baseline wander and intermittent noise.”

Cause

What was observed during troubleshooting.

Example:
Electrodes were partially detached, and the ECG cable showed minor wear. Environmental EMI from nearby infusion pumps was also present.

Resolution

What action was taken.

Example:
Replaced electrodes, secured ECG cable connections, minimized nearby EMI. Device self-test passed after corrections.

Helpful Details to Include

Final Thought

Patient safety and accurate monitoring are paramount. Logical, stepwise troubleshooting prevents unnecessary internal repairs and ensures the device is safe for use. Proper documentation helps track recurring issues and supports cost-effective maintenance.

That is successful troubleshooting.