Physio-Control LIFEPAK 20 / 20e

Error: 9007 – Shock Not Delivered

On this page

Asset Type

Defibrillator

Manufacturer

Physio-Control

Model

LIFEPAK 20 / 20e

What This Guide Helps With

Troubleshooting Error 9007 when the LIFEPAK 20 / 20e reports that a requested defibrillation shock was not delivered, potentially preventing critical therapy delivery.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Check the Power Source and Device Condition

4. Inspect Therapy Cables and Connections

5. Verify the Controlled Test Setup

6. Isolate External Accessories

7. Reload the Device Software

8. Perform the Required Performance Inspection Procedure

If the Problem Persists

If Error 9007 remains after confirming the reported error, checking stable power, inspecting therapy cables and connections, verifying the controlled test setup, isolating external accessories, and reloading the device software, then external and common causes have been ruled out.

A persistent Shock Not Delivered condition strongly indicates a device-level therapy fault requiring further authorized repair or bench evaluation.

Knowing when to stop is proper troubleshooting. A defibrillator with unresolved shock-delivery failure should never be returned to service based only on an error clearing temporarily.

Clinical Use Tip

Never troubleshoot Error 9007 on an active patient. Move the patient to a backup verified defibrillator first and maintain therapy continuity throughout the evaluation.

A prior Shock Not Delivered event should be treated seriously even if the device later appears functional. Complete corrective action and the required PIP before clinical release.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"LIFEPAK 20 / 20e reported Service Log Error 9007 – Shock Not Delivered. Device was removed from service due to reported failure of defibrillation therapy delivery."

Cause

What was observed during troubleshooting.

Example:
"External power, therapy connections, test setup, and accessible accessories were evaluated with no external cause identified. Error condition required manual-specified software corrective action."

Resolution

What action was taken.

Example:
"Reloaded device software using approved service process. Completed required PIP and controlled defibrillator therapy testing. Verified charging, shock delivery, and applicable delivered-energy performance passed before return to service."

Helpful Details to Include (If Known)

Final Thought

Error 9007 is a therapy-critical condition because it records that a shock was not delivered. Clinical Engineering troubleshooting should begin with patient safety, confirm the actual Service Log event, rule out external setup and accessory causes, and then follow the specified software corrective action.

The device should return to service only after the required PIP confirms reliable therapy performance. Clear documentation of the original failure, corrective action, testing, and final disposition is essential. Safe troubleshooting depends on a logical external-first process, appropriate escalation, and complete documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide