Puritan Bennett 560

Battery Not Charging or Shortened Runtime

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

560

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting Puritan Bennett 560 ventilators experiencing battery issues, such as failure to charge or shortened runtime. These problems can compromise backup power during transport or power outages. Most causes are external or easily verifiable, including power supply problems, battery seating, or degraded battery health. This guide focuses on safe, step-by-step troubleshooting before escalating for repair.

Step-by-Step Troubleshooting

Verify AC Power Source

Inspect Power Cord and Connections

Check Battery Seating

Observe Ventilator Indicators

Test with Spare/Service Battery

Allow Full Charge Cycle

If the Problem Persists

Common external causes have been ruled out. The battery is likely internally degraded or the charging circuit may be faulty.

Clinical Use Tip

Never rely on a suspected faulty battery during patient transport. Always move the patient to another ventilator with verified battery functionality before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator battery not charging and provides only 10 minutes of runtime.”

Cause

What was observed during troubleshooting.

Example:
Battery seated correctly but fails to accept charge; LED indicates low battery despite AC power.

Resolution

What action was taken.

Example:
Replaced battery with known good unit; device charges and holds runtime as expected. Original battery labeled for replacement/repair.

Helpful Details to Include

Final Thought

Patient safety comes first. Logical, stepwise troubleshooting can quickly identify external causes, but knowing when to escalate ensures safe operation. Proper documentation via CCR protects both the patient and the department.

That is successful troubleshooting.