Puritan Bennett 560

Circuit Disconnect Alarms

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

560

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting Puritan Bennett 560 ventilators that trigger circuit disconnect alarms. These alarms indicate that the ventilator senses a break or leak in the breathing circuit, which can compromise patient ventilation. Most causes are external or easily verifiable, such as loose tubing, disconnections, or obstructions, before considering internal device faults.

Step-by-Step Troubleshooting

Verify Patient Safety First

Check Circuit Connections

Inspect for Leaks or Damage

Verify Flow Sensors and Adapters

Check Ventilator Settings

Power Cycle the Ventilator

Observe Alarm Behavior

If the Problem Persists

If all external checks are complete and the alarms continue, the issue may be internal (pressure sensors or flow circuitry).

Action:

Knowing when to escalate is part of proper troubleshooting.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
User reports repeated “circuit disconnect” alarms while ventilating patient.

Cause

What was observed during troubleshooting.

Example:
Loose expiratory limb connection and kinked inspiratory tubing observed during inspection.

Resolution

What action was taken.

Example:
Reconnected tubing securely, replaced damaged sections, verified circuit integrity, ventilator returned to service.

Helpful Details to Include

Final Thought

Proper troubleshooting prioritizes patient safety, starts with simple external checks, and avoids unnecessary internal interventions. Documenting actions clearly ensures continuity of care and supports cost-effective maintenance.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide