Puritan Bennett 560

Turbine Failure or Insufficient Flow

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

560

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting Puritan Bennett 560 ventilators experiencing turbine failure or insufficient airflow. These issues can lead to inadequate ventilation, trigger alarms, or compromise patient safety. Most causes are external or easily verifiable, such as blocked airways, power issues, or environmental factors, before considering internal turbine failure.

Step-by-Step Troubleshooting

Check Power Supply

Inspect Air Intake and Filters

Check Circuit and Tubing Connections

Observe Turbine Operation

Run Ventilator Self-Test or Diagnostic Mode

Check Environmental Conditions

If the Problem Persists

Common external causes have been ruled out. The turbine is likely mechanically or electronically failing.

Action:

Knowing when to escalate is part of proper troubleshooting.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator delivering insufficient flow; alarms triggered; turbine suspected.”

Cause

What was observed during troubleshooting.

Example:
“Air intake filter partially blocked; tubing unobstructed; turbine self-test indicates error code 12 (turbine malfunction).”

Resolution

What action was taken.

Example:
“Replaced air filter and ran diagnostic; issue persists. Device labeled Out of Service and sent for repair.”

Helpful Details to Include

Final Thought

Patient safety is paramount. Logical, stepwise troubleshooting prevents unnecessary internal damage and ensures only qualified personnel handle potential turbine failures. Proper documentation ensures traceability and reinforces the value of Clinical Engineering in maintaining safe ventilation devices.

That is successful troubleshooting.