Puritan Bennett 840

Alarm: INSPIRATION TOO LONG prolonged inspiratory time

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

840

What This Guide Helps With

This guide helps troubleshoot the Puritan Bennett 840 Ventilator alarm INSPIRATION TOO LONG, which indicates prolonged inspiratory time causing improper breath cycling, potential air trapping, or unsafe ventilation timing. This condition can affect normal exhalation and may compromise safe ventilation if not addressed promptly.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Patient Circuit for Obstruction or Flow Restriction

Check Patient Condition and Interaction with the Ventilator

Verify Ventilator Settings

Check Flow Sensors and Circuit Components

Run Ventilator Self Tests

Stop When Resolved

If the Problem Persists

If the INSPIRATION TOO LONG alarm continues after circuit obstruction, settings, patient interaction, flow sensor condition, and external component issues have been ruled out, remove the ventilator from service.

Label the device Out of Service and send it for repair or bench evaluation. Persistent timing-related alarms after external checks strongly suggest an internal control or sensing failure.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to backup ventilation first before making adjustments or continuing diagnostic steps.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming 'INSPIRATION TOO LONG' during use with prolonged inspiratory phase observed."

Cause

What was observed during troubleshooting.

Example:
"Restricted circuit airflow and/or inspiratory time setting exceeded expected limits, causing delayed cycling to exhalation."

Resolution

What action was taken.

Example:
"Cleared circuit obstruction, verified settings, cleaned and reseated flow sensor, and ran SST/EST. Alarm cleared and unit returned to service."

Helpful Details to Include (If Known)

Final Thought

Alarms like INSPIRATION TOO LONG reflect real ventilation timing problems that should be approached with a logical external-first process. Start with patient safety, confirm the behavior, rule out circuit and setting issues, and escalate when appropriate. Clear documentation supports safe follow-up and repair decisions.

That is successful troubleshooting.