Puritan Bennett 840

Alarm: LOW INSP PRESSURE Inspiratory pressure below expected target

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

840

What This Guide Helps With

LOW INSP PRESSURE alarm indicating insufficient inspiratory pressure, resulting in inadequate ventilation or failure to reach set tidal volume.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Alarm and Behavior

3. Check Patient Condition

4. Check Circuit Connections

5. Check for Leaks

6. Verify Ventilator Settings

7. Inspect Sensors and Circuit Components

8. Run Extended Self-Test (EST)

If the Problem Persists

If all external causes have been ruled out and the alarm continues, internal pressure regulation, sensors, or control systems may be failing. Remove the ventilator from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot a ventilator while actively supporting a patient. Always move the patient to a backup device before performing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming LOW INSP PRESSURE with reduced tidal volume during use."

Cause

What was observed during troubleshooting.

Example:
"Leak identified in patient circuit preventing adequate inspiratory pressure."

Resolution

What action was taken.

Example:
"Reseated circuit connections, eliminated leak, verified normal pressure delivery and alarm cleared."

Helpful Details to Include (If Known)

Final Thought

LOW INSP PRESSURE alarms are most commonly caused by leaks, patient condition changes, or incorrect settings. Always rule out external causes first and escalate only after confirming system integrity.

That is successful troubleshooting.