Puritan Bennett 840

Alarm: ↓O2% Measured O2 below set value

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

840

What This Guide Helps With

This guide addresses the ↓O2% alarm indicating delivered oxygen concentration is below the set value, which can lead to inadequate oxygenation and increased patient risk.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Oxygen Supply Source

Inspect Oxygen Inlet Components

Perform O₂ Sensor Calibration

Run Extended Self-Test (EST)

Check Oxygen PSOL Function

Replace Oxygen Sensor (If Needed)

If the Problem Persists

If the ↓O2% alarm continues after completing all external checks, internal components such as the PSOL or control system may be at fault. Remove the device from service, label it Out of Service, and send it for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot oxygen delivery issues on an active patient. Always transition the patient to a backup ventilator or manual ventilation before continuing diagnostics.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming ↓O2% with delivered FiO₂ lower than set value during use."

Cause

What was observed during troubleshooting.

Example:
"Oxygen sensor failed calibration and showed inaccurate low FiO₂ readings."

Resolution

What action was taken.

Example:
"Replaced oxygen sensor, recalibrated system, verified accurate FiO₂ delivery, and confirmed alarm cleared."

Helpful Details to Include (If Known)

Final Thought

Low oxygen alarms directly impact patient safety and must be addressed immediately. Always verify supply, calibration, and measurement accuracy before escalating to internal component failure.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide