Puritan Bennett 840

Alarm: O2 SENSOR Oxygen sensor out of calibration or failed

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

840

What This Guide Helps With

This guide addresses the Alarm: O2 SENSOR Oxygen sensor out of calibration or failed, which indicates inaccurate oxygen readings or loss of FiO₂ monitoring. This condition can impact safe ventilation delivery and patient oxygen management.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Perform O2 Sensor Calibration (Most Common Cause)

4. Check Oxygen Supply Source and Stability

5. Inspect Sensor Connection and Physical Condition

6. Determine if Sensor Has Reached End of Life

7. Optional: Disable O2 Sensor (If Clinically Appropriate)

If the Problem Persists

If calibration fails, connections are verified, and a known-good sensor does not resolve the issue, internal failure is likely.

Remove the ventilator from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot on an active patient. Always move the patient to a backup ventilator before performing troubleshooting steps.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator displaying O2 SENSOR alarm with inaccurate FiO₂ readings during operation."

Cause

What was observed during troubleshooting.

Example:
"Oxygen sensor failed calibration and showed unstable readings, indicating sensor degradation."

Resolution

What action was taken.

Example:
"Replaced oxygen sensor, completed successful calibration, and verified accurate FiO₂ readings."

Helpful Details to Include (If Known)

Final Thought

Oxygen sensor alarms directly impact patient safety and monitoring accuracy. Always prioritize calibration and external checks first, but recognize when the sensor must be replaced or the device escalated for repair. Clear documentation ensures reliability and traceability.

That is successful troubleshooting.