Puritan Bennett 840

Alarm: ↑PPEAK Peak airway pressure exceeds limit

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

840

What This Guide Helps With

Alarm: ↑PPEAK indicating high airway pressure, which can lead to barotrauma, inadequate ventilation delivery, or ventilator breath termination due to pressure limits being exceeded.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Patient Condition

Inspect the Endotracheal (ET) Tube

Check the Patient Circuit for Obstruction

Review Ventilator Settings

Run Ventilator Self-Test (SST/EST) if Needed

If the Problem Persists

If all external causes (patient condition, ET tube, circuit, and settings) have been ruled out and the ↑PPEAK alarm continues:

Clinical Use Tip

Never troubleshoot high airway pressure alarms on an active patient. Always secure ventilation with a backup method before investigating the device.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming ↑PPEAK with high airway pressures during use; intermittent breath termination observed."

Cause

What was observed during troubleshooting.

Example:
"Kinked expiratory limb of circuit causing increased resistance and elevated peak airway pressure."

Resolution

What action was taken.

Example:
"Adjusted/replaced circuit tubing, confirmed normal airway pressures, alarm cleared, ventilator returned to service."

Helpful Details to Include (If Known)

Final Thought

High airway pressure alarms must be treated as both patient and equipment concerns. Always start with the patient, then work outward through the airway and circuit. Escalate only after all external causes are eliminated and document clearly.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide