Puritan Bennett 840

Alarm: ↑PVENT Inspiratory pressure exceeds 100 cmH₂O

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

840

What This Guide Helps With

The ↑PVENT alarm indicates inspiratory pressure has exceeded 100 cmH₂O. This is a critical high-pressure condition that can prevent proper ventilation and place the patient at risk for barotrauma, lung injury, or inadequate breath delivery.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Alarm Condition

3. Check Patient Condition

4. Inspect the Breathing Circuit

5. Review Ventilator Settings

6. Perform System Self Test (SST)

7. Perform Extended Self Test (EST)

8. Swap External Components

9. Stop When Resolved

If the Problem Persists

If the ↑PVENT alarm continues after all external checks and testing, internal pressure sensing or regulation failure is likely.

Clinical Use Tip

Never troubleshoot a ventilator experiencing high-pressure alarms on an active patient. Always transition the patient to a backup ventilation method before continuing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming ↑PVENT with high inspiratory pressure and low volume delivery."

Cause

What was observed during troubleshooting.

Example:
"Kinked breathing circuit causing increased resistance and elevated pressure."

Resolution

What action was taken.

Example:
"Replaced circuit and verified normal pressures. Completed SST successfully."

Helpful Details to Include (If Known)

Final Thought

High inspiratory pressure alarms must be addressed immediately. Always rule out patient and circuit causes first, and escalate to device service when internal issues are suspected.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide