Puritan Bennett 840

Alarm: ↓VTE MAND Mandatory tidal volume below limit

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

840

What This Guide Helps With

↓VTE MAND alarm indicating delivered mandatory tidal volume is too low, resulting in inadequate ventilation to the patient.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Patient Condition (Most Critical First Step)

4. Inspect Patient Circuit for Leaks or Disconnections

5. Check Ventilator Settings

6. Run Short Self Test (SST)

7. Stop When Resolved

If the Problem Persists

If the ↓VTE MAND alarm continues after external causes are ruled out, this indicates a ventilator performance issue such as flow delivery, sensors, or internal control fault.

Clinical Use Tip

Never troubleshoot low tidal volume alarms on an active patient without backup ventilation available. Always transition the patient first before continuing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ventilator alarming ↓VTE MAND with low delivered tidal volume during mandatory breaths."

Cause

What was observed during troubleshooting.

Example:
"Unable to restore proper tidal volume after verifying patient airway, circuit integrity, and ventilator settings; SST performed—suspected internal ventilator issue."

Resolution

What action was taken.

Example:
"Removed ventilator from service, labeled Out of Service, and sent to Biomed shop/vendor for further evaluation."

Helpful Details to Include (If Known)

Final Thought

Low tidal volume alarms require immediate attention. Always start with patient safety and external checks, but recognize when persistent issues indicate deeper device faults requiring escalation.

That is successful troubleshooting.