Puritan Bennett 840

Flow Sensor Errors or Calibration Failure

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

840

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting Puritan Bennett 840 ventilators that display flow sensor errors or fail calibration. Flow sensor issues can prevent accurate ventilation, trigger alarms, or compromise patient safety. Most causes are external, such as sensor obstruction, improper setup, or environmental conditions. This guide emphasizes safe, step-by-step troubleshooting before assuming internal sensor or board failure.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Check Power and Connections

Inspect the Flow Sensor

Verify Circuit Assembly

Perform Calibration

Check for Software or Settings Issues

If the Problem Persists

External and setup-related causes have been ruled out. The flow sensor or internal electronics may be defective.

Remember: knowing when to escalate is proper troubleshooting.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator displays flow sensor error and fails calibration.”

Cause

What was observed during troubleshooting.

Example:
Observed debris in flow sensor and/or misaligned circuit connections preventing calibration.

Resolution

What action was taken.

Example:
Cleaned or replaced flow sensor, reattached circuits properly, recalibrated ventilator. If errors persisted, removed ventilator from service and sent for repair.

Helpful Details to Include

Final Thought

Patient safety and logical troubleshooting are paramount. Always start with external checks and simple fixes before assuming internal failure. Proper documentation ensures maintenance history and safe equipment management. Escalating promptly when needed protects both patients and staff.

That is successful troubleshooting.