Puritan Bennett 840

Internal Battery Degradation

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

840

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting and managing internal battery degradation in Puritan Bennett 840 ventilators. Battery degradation can lead to shortened backup time, inability to hold charge, or unexpected power loss during transport or power interruptions. Most causes are related to normal battery wear over time, but proper verification and safe handling are essential before escalation.

Step-by-Step Troubleshooting

Verify Power Source

Check Battery Charge Status

Inspect Battery Connection

Perform Battery Test (if available)

Observe Ventilator During AC Power Loss Simulation

If the Problem Persists

If battery fails to hold charge, shows rapid depletion, or fails the internal test:

Clinical Engineering Note: Knowing when to escalate is proper troubleshooting. Internal battery replacement requires manufacturer-approved service.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator battery backup only lasts ~10 minutes and battery warning displayed.”

Cause

What was observed during troubleshooting.

Example:
“Observed internal battery capacity below manufacturer specifications; battery fails internal test and cannot maintain power during AC disconnect simulation.”

Resolution

What action was taken.

Example:
“Ventilator removed from service, labeled Out of Service, and sent for authorized battery replacement.”

Helpful Details to Include

Final Thought

Proper identification of internal battery degradation is critical for patient safety. Troubleshooting should progress from simple verification of AC power and connections to testing battery performance. Escalating to manufacturer service for replacement avoids unnecessary risk and ensures reliable ventilator operation. Accurate CCR documentation supports accountability and departmental records.

That is successful troubleshooting.