Puritan Bennett 980

Air Supply, Compressor, or Gas Source Failure

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

980

What This Guide Helps With

Troubleshooting air-supply alarms, compressor failure, reduced gas pressure, disconnected hoses, closed outlets, restricted fittings, or unavailable wall-gas sources.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a gas-source or compressor failure while the Puritan Bennett 980 is actively supporting a patient.

Expected outcome: The patient is safely supported without relying on a ventilator with an unreliable gas source.

Continue Clinical Engineering troubleshooting only after the affected PB980 has been removed from patient use. The PB980 is intended for neonatal, pediatric, and adult critical-care ventilation and should be operated by trained healthcare professionals.

2. Confirm the Exact Failure

Review the active alarm, alarm history, and reported behavior. Determine whether the issue involves:

Record the exact alarm wording and any diagnostic code before cycling power.

Expected outcome: The failure is clearly identified as an air-source, compressor, connection, or ventilator-related problem.

3. Verify the Facility Gas Source

Confirm that the medical-air wall outlet or approved external gas source is available.

Expected outcome: A functioning facility medical-air source is confirmed.

If the wall source is unavailable, relocate the ventilator to a verified outlet or address the facility gas issue. Stop if normal operation is restored.

4. Inspect the Air-Supply Hose

Inspect the entire medical-air hose from the wall outlet or compressor to the ventilator.

Check for:

Reconnect each fitting securely without forcing incompatible connectors.

Expected outcome: The hose is correctly connected, unrestricted, undamaged, and appropriate for medical air.

If correcting the hose connection clears the alarm and the ventilator passes testing, stop.

5. Test the Wall Outlet and Hose Independently

Use an approved medical-gas pressure tester or another facility-approved method to verify the wall outlet.

When permitted:

Do not vent medical gas unnecessarily or use unapproved gauges and adapters.

Expected outcome: The fault is isolated to the wall outlet, gas hose, or ventilator.

Replace a defective hose or refer an outlet problem to Facilities. Stop if the PB980 operates normally afterward.

6. Check the Ventilator Air-Inlet Connection

Inspect the PB980 air-inlet fitting externally.

Look for:

Do not insert tools into the inlet or attempt internal fitting repair during floor troubleshooting.

Expected outcome: The external air-inlet connection is clean, secure, and free of visible damage or leakage.

7. Evaluate the Compressor Configuration

Determine whether the PB980 is expected to use:

Confirm that the compressor is connected, powered, and configured according to the facility’s approved setup.

For an external compressor, check:

Do not remove compressor covers or bypass protective devices.

Expected outcome: The compressor powers on, produces stable air pressure, and shows no external fault condition.

8. Confirm Adequate AC Power for the Compressor

If a compressor is used, connect it directly to a verified hospital-grade receptacle.

Expected outcome: The compressor receives stable AC power and remains operating.

A repeated trip, burning smell, excessive heat, or immediate shutdown requires removal from service.

9. Check Accessible Intake and Cooling Areas

Inspect externally accessible compressor and ventilator air intakes.

Expected outcome: Air intake and cooling paths are externally clear.

If clearing an obstruction restores stable compressor operation, continue with functional testing before returning the system to use.

10. Restart the System Off-Patient

After confirming the external gas source, hoses, fittings, and compressor condition:

Do not repeatedly restart a ventilator that continues to report a gas-delivery failure.

Expected outcome: The PB980 completes startup without recurring air-supply or compressor alarms.

11. Run the Required Pre-Use Test

Connect an approved test circuit and test lung. Complete the manufacturer-required startup or short self-test according to facility procedure and the current operator’s manual.

Confirm:

The manufacturer identifies the operator’s manual as the primary operating and troubleshooting reference for the PB980.

Expected outcome: The ventilator passes the required pre-use testing with a verified gas source.

If the ventilator passes after replacing a hose or correcting the external source, document the repair and return it according to facility policy.

12. Compare With a Known-Good Gas Source

When the failure remains unclear, connect the PB980 to a known-good compatible air source using a known-good hose.

Expected outcome: The comparison confirms whether the failure follows the source equipment or remains with the PB980.

If the alarm remains with verified gas pressure and a known-good hose, suspect an internal pressure-regulation, gas-mixing, sensor, inlet, or compressor-system problem.

13. Review Logs and Diagnostic Information

Record available information before removing the unit from service:

Do not clear logs before the repair technician has reviewed them.

Expected outcome: The repair team receives enough information to reproduce and diagnose the failure efficiently.

If the Problem Persists

If the Puritan Bennett 980 continues to report an air-supply, compressor, or gas-source failure after the facility source, hoses, connections, electrical supply, accessible filters, and compressor setup have been verified, common external causes have been ruled out.

The failure may involve an internal gas-pressure sensor, regulator, inlet assembly, compressor component, gas-delivery subsystem, wiring connection, or control circuit.

The ventilator should be:

Do not bypass gas-source alarms, internal sensors, regulators, or compressor safety devices. Knowing when to stop external troubleshooting is proper Clinical Engineering practice.

Clinical Use Tip

Never troubleshoot a PB980 gas-source failure while it remains connected to an active patient. Move the patient to a verified ventilator first and confirm effective ventilation and alarms before evaluating the affected unit.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Puritan Bennett 980 displayed an air-supply alarm and could not maintain normal operation using the room’s medical-air outlet."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the air hose quick-connect fitting was not fully seated, causing intermittent loss of inlet pressure."

Resolution

What action was taken.

Example:
"Reseated and inspected the air hose, verified the wall outlet, completed the ventilator pre-use test with a test lung, and confirmed that the alarm did not recur."

Helpful Details to Include (If Known)

Final Thought

Air-supply failures require immediate attention because stable gas delivery is essential to safe ventilation. Begin with patient safety, verify the facility source and connections, test one component at a time, and escalate when the failure remains with the ventilator. Clear CCR documentation preserves the evidence needed for efficient repair.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide