Puritan Bennett 980

Alarm Speaker or Audible Alarm Failure

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

980

What This Guide Helps With

Troubleshooting absent, weak, distorted, intermittent, or unexpected alarm audio caused by volume settings, audio pause, obstructions, software state, or speaker failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot an audible alarm failure while the Puritan Bennett 980 is actively supporting a patient.

Expected outcome: The patient is safely supported by a ventilator with verified alarm operation.

Continue Clinical Engineering troubleshooting only after the affected PB980 has been removed from patient use.

2. Confirm the Exact Audio Failure

Determine whether the issue involves:

Observe the visual alarm indicators and record any displayed technical alarm, diagnostic message, or system fault.

Expected outcome: The failure is clearly defined as a complete, intermittent, volume-related, or alarm-specific problem.

3. Check for an Active Audio Pause

Verify whether the audio pause function is active.

Some PB980 alarms may remain silent during an active audio-pause period, depending on alarm behavior and priority.

Expected outcome: Alarm audio returns after the pause is canceled or expires.

If normal audio returns, document the operating condition and stop troubleshooting.

4. Verify the Alarm Volume Setting

Use the alarm volume control to confirm that the volume is not set too low for the environment.

The PB980 generates a sample tone during alarm-volume adjustment. Its alarm volume may also increase automatically when a high-priority alarm remains unacknowledged.

Expected outcome: A clear sample tone is heard and the alarm volume responds to adjustment.

If the sample tone is absent, distorted, or intermittent, continue troubleshooting.

5. Inspect the Speaker Opening and Ventilator Exterior

Inspect the ventilator externally for conditions that could reduce audible output.

Do not insert tools, compressed air, or cleaning fluid into speaker openings.

Expected outcome: The speaker path is unobstructed and no external damage is found.

If an obstruction caused the issue, remove it safely, clean according to facility procedures, retest, and stop if alarm audio is restored.

6. Check External Connections and Accessories

Inspect externally connected equipment that could be associated with the report.

Expected outcome: The local audible alarm is evaluated independently from external alarm accessories.

If disconnecting a damaged or incompatible accessory restores normal operation, replace or evaluate that accessory before returning the ventilator to service.

7. Perform a Controlled Power Restart

With the ventilator removed from patient use:

Do not repeatedly cycle power if the ventilator displays a persistent technical fault or fails its startup checks.

Expected outcome: The ventilator starts normally and audible indications return without system faults.

If the restart restores normal audio, continue with a complete alarm verification before returning the device to service.

8. Test Alarm Audio Using Approved Test Equipment

Connect the PB980 to the required gas supplies, an approved patient circuit, and a test lung.

Create alarm conditions only through the facility-approved functional test or manufacturer alarm-test procedure. Verify:

The manufacturer specifies periodic alarm testing and identifies test-lung, patient-circuit, stable AC-power, air, and oxygen requirements for the alarm test.

Expected outcome: Each tested alarm produces a clear and repeatable audible indication corresponding with its visual alarm.

If an alarm does not annunciate correctly, confirm the test setup and ventilator settings once, then repeat the test. Do not return the ventilator to service if the failure remains.

9. Compare the Report With Alarm History and Technical Messages

Review available alarm history or diagnostic information without altering protected service settings.

Look for:

Expected outcome: Available history either supports an intermittent failure or confirms that no external operating condition explains the complaint.

Do not clear useful logs before documenting or escalating the failure.

10. Complete a Final Functional Verification

Before considering the ventilator operational:

Expected outcome: Alarm audio is clear, repeatable, and fully functional during controlled testing.

If any alarm speaker or audible-annunciation failure remains intermittent or unresolved, remove the ventilator from service.

If the Problem Persists

If audio pause, volume settings, speaker obstruction, accessories, power state, test setup, and external conditions have been ruled out, the failure is likely internal.

Possible internal causes may include the speaker assembly, audio circuitry, internal wiring, GUI electronics, software, or system communication hardware. Do not proceed with deep internal disassembly unless authorized, trained, and following current manufacturer service documentation.

The ventilator should be:

A ventilator must not be returned to clinical use when local audible alarm operation is absent, distorted, intermittent, or unverified. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot an alarm speaker problem on an active patient. Move the patient to another verified ventilator first and confirm that its visual and audible alarms operate correctly.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Puritan Bennett 980 displayed visual alarms but produced no audible alarm during operation."

Cause

What was observed during troubleshooting.

Example:
"Alarm volume and audio-pause settings were verified, external obstructions and accessories were ruled out, and no sample tone or audible output occurred during controlled alarm testing."

Resolution

What action was taken.

Example:
"Removed the ventilator from service, labeled it Out of Service, documented the failed alarm test, and sent the unit for authorized speaker and audio-system evaluation."

Helpful Details to Include (If Known)

Final Thought

Reliable audible alarms are essential to ventilator safety. A logical evaluation of settings, external conditions, accessories, startup behavior, and controlled alarm performance helps identify correctable causes while ensuring that unresolved failures are escalated appropriately. Complete CCR documentation preserves the clinical complaint, troubleshooting findings, and final disposition.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide