On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting battery charging failures, reduced runtime, unexpected battery depletion, AC power recognition problems, loose connections, aging batteries, or internal charging-system faults.
Step-by-Step Troubleshooting
1. Ensure Patient Safety First
Do not troubleshoot a battery or charging problem while the Puritan Bennett 980 is actively supporting a patient.
- Notify respiratory therapy and the clinical team that backup battery operation may be unavailable or unreliable.
- Transfer the patient to another verified ventilator using the facility’s approved clinical procedure.
- Confirm that the replacement ventilator is operating correctly and has adequate backup power.
- Do not intentionally disconnect AC power while the affected ventilator is supporting a patient.
Expected outcome: The patient is no longer dependent on a ventilator with uncertain backup power.
Continue Clinical Engineering troubleshooting only after the affected PB980 has been safely removed from patient use. The PB980 is intended for trained healthcare professionals in acute or critical-care environments, making dependable power operation a patient-safety requirement.
2. Confirm the Reported Battery Problem
Determine whether the issue involves:
- Battery not charging
- Battery charge percentage not increasing
- Battery indicator missing or showing an error
- Unexpected switch to battery while connected to AC
- Battery discharging unusually quickly
- Ventilator shutting down shortly after AC power is removed
- Failure involving the primary battery, extended battery, or both
Record any displayed alarms, status messages, battery indicators, and the approximate runtime reported by staff.
Expected outcome: The exact failure pattern is identified before parts are replaced.
If the report cannot be reproduced and all power indications remain normal, document the findings and continue with an operational battery check.
3. Verify the AC Power Source
Connect the ventilator directly to a known-good hospital-grade receptacle.
- Avoid extension cords, adapters, switched power strips, and unverified isolation-power outlets.
- Test the original receptacle with an approved outlet tester or another known-good device.
- Confirm the receptacle provides stable power and is not controlled by a wall switch.
- Check whether nearby equipment also experienced a power interruption.
Expected outcome: The ventilator recognizes AC power and the battery begins or continues charging.
If the battery charges normally from the known-good receptacle, the original outlet or external power source was the likely cause. Stop troubleshooting and address the facility power issue.
4. Inspect the Power Cord and Connections
Disconnect the ventilator from AC power before inspecting the cord.
Check for:
- Loose connection at the wall receptacle
- Incomplete connection at the ventilator
- Bent, recessed, contaminated, or damaged connector contacts
- Cuts, crushing, exposed conductors, discoloration, or heat damage
- A loose plug that falls out of the receptacle
- Evidence that the cord has been stretched or repeatedly rolled over
Reconnect the cord firmly at both ends. When approved by facility policy, test with a known-good compatible power cord.
Expected outcome: AC power remains continuously recognized without movement-related interruptions.
If replacing or reseating the cord restores charging, remove the defective cord from service and stop troubleshooting.
5. Confirm AC and Battery Status Indications
Power the ventilator from the verified AC source and observe the graphical user interface and status display.
Confirm that:
- AC power is recognized
- The battery is detected
- A charging indication is present when charging is required
- No battery, power-system, or charging-related alarm remains active
- The charge level does not continue decreasing while connected to AC
Allow enough time to determine whether the displayed charge level is increasing. A severely depleted battery may not show an immediate large percentage change.
Expected outcome: The PB980 detects both AC power and the installed battery, and the charge level begins trending upward.
If AC power is not recognized despite a verified outlet and cord, stop external troubleshooting and escalate for bench evaluation.
6. Inspect Accessible Battery Installation
With the ventilator powered down and removed from AC power, inspect the externally accessible battery installation according to facility procedures.
Check for:
- Battery not fully seated or latched
- Loose accessible battery connection
- Incorrect battery installation
- Damaged latch, cover, housing, or connector area
- Swelling, cracking, leakage, corrosion, or discoloration
- Unusual heat or chemical odor
- Missing battery where one is expected
Do not use a swollen, leaking, overheated, cracked, or physically damaged battery.
Expected outcome: The battery is correctly installed, mechanically secure, and free from visible damage.
If reseating an undamaged battery restores detection and charging, complete the required operational verification and stop troubleshooting.
7. Check Battery History and Condition
Review the equipment history, battery label, installation date, previous work orders, and recent runtime complaints.
Look for:
- Battery age beyond the facility’s approved replacement interval
- Long periods of storage without charging
- Repeated deep discharge events
- Previous battery alarms
- Gradual runtime reduction
- A recent battery replacement followed by continued charging problems
Battery capacity decreases with age and use. A battery may appear fully charged but lose voltage rapidly when placed under load.
Expected outcome: Battery age and service history are consistent with acceptable performance.
If the battery is aged, has repeated runtime complaints, or has exceeded the approved replacement interval, replace it with the correct manufacturer-approved battery and complete post-replacement testing.
8. Compare Primary and Extended Battery Behavior
When the ventilator is equipped with more than one battery, review each battery’s status separately if the interface permits.
Determine whether:
- One battery is detected and the other is missing
- One battery charges while the other does not
- Both batteries show the same abnormal condition
- Runtime improves when a known-good battery is installed in the affected position
Do not interchange batteries or positions unless permitted by the applicable service documentation and facility procedure.
Expected outcome: The problem is isolated to a specific battery or shown to affect the ventilator’s complete charging system.
If one battery consistently fails while another approved battery functions correctly, replace the defective battery and stop troubleshooting after successful verification.
9. Perform a Controlled Charging Check
Connect the PB980 to verified AC power in a safe bench area.
- Record the starting battery percentage or status.
- Allow the ventilator to remain connected for an appropriate charging period.
- Confirm that AC recognition remains stable.
- Record whether the battery indication increases.
- Monitor for battery alarms, abnormal heat, smell, or repeated charging interruptions.
Do not cover ventilation openings or leave a battery showing abnormal heat unattended.
Expected outcome: Battery charge increases without alarms, excessive heat, or loss of AC recognition.
If the battery remains unchanged or continues discharging while AC is continuously recognized, suspect a failed battery or internal charging-system problem.
10. Perform a Controlled Runtime Check
Perform this check only off-patient in a controlled environment and according to facility policy.
- Begin with a fully charged battery indication.
- Configure the ventilator with an approved test circuit and test lung.
- Record the ventilation settings and installed accessories.
- Disconnect AC power while maintaining close observation.
- Confirm that the ventilator transfers to battery without interruption.
- Monitor battery status, alarms, and elapsed operating time.
- Reconnect AC power before battery exhaustion or at the facility’s defined test endpoint.
Runtime depends on battery condition, ventilation settings, accessories, battery configuration, and environmental conditions. Compare the result with the applicable manufacturer specification and facility acceptance criteria rather than relying only on staff estimates.
Expected outcome: The ventilator transfers correctly to battery and meets the approved runtime requirement.
If the battery depletes prematurely, replace the suspected battery with an approved known-good battery and repeat the test.
11. Test With a Known-Good Approved Battery
When authorized, install a known-good compatible battery with verified service history.
- Confirm correct seating and detection.
- Connect the ventilator to AC and verify charging.
- Complete a controlled battery transfer and runtime check.
- Compare the result with the original battery.
Expected outcome: A known-good battery charges and provides acceptable runtime.
If the known-good battery resolves the issue, replace the original battery and complete all required performance and safety verification.
If the known-good battery also fails to charge or provides abnormal operation, suspect the ventilator’s internal charger, battery communication, power-distribution circuitry, wiring, or related electronics.
12. Complete Final Operational Verification
After correcting the problem:
- Confirm stable AC operation.
- Confirm the battery is detected and charging.
- Confirm transfer from AC to battery occurs without ventilation interruption.
- Confirm return to AC power is recognized.
- Verify battery and power alarms using the approved test procedure.
- Confirm no unusual heat, odor, or noise is present.
- Verify the ventilator passes the facility’s required functional and electrical-safety testing.
- Confirm the battery is adequately charged before returning the ventilator to clinical service.
Expected outcome: The PB980 operates correctly on AC and battery power and meets the facility’s return-to-service criteria.
If all checks pass, return the ventilator to service and document the completed work.
If the Problem Persists
If the PB980 still does not detect or charge an approved known-good battery, loses AC recognition, repeatedly switches power sources, or fails the runtime test, common external causes have been ruled out.
The failure may involve the internal charging circuit, power supply, battery interface, internal wiring, communication circuitry, or power-distribution electronics.
The ventilator should be:
- Removed from service
- Labeled Out of Service
- Sent for qualified repair or bench evaluation
- Evaluated using the applicable manufacturer service documentation
- Subjected to complete performance and electrical-safety verification before return to use
Do not continue repeatedly discharging batteries or operating the ventilator with uncertain backup power. Knowing when to stop and escalate is proper troubleshooting.
Clinical Use Tip
Never troubleshoot questionable battery performance on an active patient. Move the patient to another verified ventilator first, and remember that AC operation alone does not prove that emergency backup power is dependable.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Respiratory therapy reported that the Puritan Bennett 980 displayed a low battery condition and operated for only several minutes after AC power was disconnected."
Cause
What was observed during troubleshooting.
Example:
"Testing found that the installed primary battery indicated a full charge but lost capacity rapidly under ventilator load; AC input and charging indications were otherwise normal."
Resolution
What action was taken.
Example:
"Replaced the failed battery with an approved replacement, verified charging, confirmed uninterrupted AC-to-battery transfer, completed a controlled runtime test, and returned the ventilator to service."
Helpful Details to Include (If Known)
- Original outlet tested
- Known-good receptacle used
- Power cord inspected
- Power cord swapped
- AC indicator behavior
- Battery charging indication
- Starting and ending battery percentage
- Primary or extended battery affected
- Battery installation date or age
- Known-good battery tested
- Runtime test settings and duration
- Battery or power alarms displayed
- Transfer to battery verified
- Return to AC power verified
- Unusual heat, swelling, leakage, or odor
- Functional verification completed
- Electrical-safety test completed
- Final device status
Final Thought
Battery troubleshooting on a ventilator requires more than confirming that the device powers on. Protect the patient first, verify external power and connections, evaluate each battery logically, and escalate when a known-good battery does not correct the failure. Clear CCR documentation preserves the evidence needed for safe equipment decisions and future trend analysis.
That is successful troubleshooting.