What This Guide Helps With

This guide addresses situations where a Puritan Bennett 980 ventilator fails its pre-use check due to a circuit leak being detected. The focus is on common, external causes such as circuit assembly, humidifier setup, and accessory connections before assuming internal pneumatic failure. Most leak failures are related to setup, not internal defects.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify Correct Circuit Type Selection

Inspect Circuit Assembly from Ventilator to Test Lung

Check the Test Lung and Port Cap

Inspect the Humidifier Setup (If Installed)

Examine the Expiratory Filter and Valve Area

Check for Cracked or Worn Tubing

Repeat Pre-Use Check After Corrections

If the Problem Persists

If a known-good circuit, test lung, filter, and proper configuration still result in a leak failure, common external causes have been ruled out.

The issue may involve:

Continuing repeated leak tests without identifying external causes is not effective troubleshooting. Knowing when to escalate protects patients and equipment.

Clinical Use Tip

Never attempt leak troubleshooting while the ventilator is supporting a patient. Transfer the patient to a confirmed functional ventilator before performing diagnostics. Leak failures can indicate compromised ventilation accuracy.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator failed pre-use check with ‘Circuit Leak Detected’ alarm in ICU prior to patient use.”

Cause

What was observed during troubleshooting.

Example:
“Loose connection identified at expiratory filter housing. Filter not fully seated, allowing leak during self-test.”

Resolution

What action was taken.

Example:
“Reseated expiratory filter, verified tight seal, reran pre-use check. Ventilator passed all tests and returned to service.”

Helpful Details to Include

Final Thought

Most ventilator leak failures are setup-related. A calm, methodical approach—checking configuration, connections, and disposables first—prevents unnecessary downtime and avoids premature escalation. Patient safety always comes first, and proper documentation ensures traceability and accountability.

That is successful troubleshooting.