Puritan Bennett 980 Ventilator

Flow Sensor Errors / Inaccurate Volumes

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

980

What This Guide Helps With

This guide addresses situations where a Puritan Bennett 980 ventilator displays flow sensor errors, inaccurate tidal volumes, volume mismatch alarms, or inconsistent measured values.

The focus is on common, external, and easily verifiable causes—such as sensor setup, circuit configuration, moisture buildup, and calibration—before assuming internal sensor or board failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Correct Circuit Is Installed

3. Inspect the Flow Sensor Placement

4. Check for Moisture or Condensation

5. Inspect the Patient Circuit for Leaks

6. Perform a System Check / Calibration

7. Replace the Flow Sensor

8. Evaluate Alarm Behavior

If the Problem Persists

If circuit is verified correct, no moisture present, no leaks detected, sensor replaced, and calibration fails or alarms persist, the issue is likely internal (flow module, sensor interface board, or internal pneumatic fault).

Knowing when to stop prevents unsafe ventilation and protects patient safety.

Clinical Use Tip

Flow accuracy directly affects delivered tidal volume and minute ventilation.

Never troubleshoot flow sensor errors on an actively ventilated patient unless they are safely transitioned to backup ventilation first.

Even small inaccuracies can be clinically significant in neonatal or low tidal volume cases.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator displaying ‘Flow Sensor Error’ and reported tidal volumes lower than set volume during use in ICU.”

Cause

What was observed during troubleshooting.

Example:
“Moisture observed inside proximal flow sensor. System check initially failed calibration. Sensor replaced and circuit drained.”

Resolution

What action was taken.

Example:
“Replaced proximal flow sensor, drained condensation from circuit, performed full system check. Calibration passed and volume readings returned to normal. Ventilator returned to service.”

Helpful Details to Include

Final Thought

Accurate flow measurement is critical to safe ventilation. Always rule out simple causes—circuit setup, moisture, leaks, and sensor condition—before assuming internal failure. Logical troubleshooting and proper escalation protect both the patient and the integrity of the equipment.

That is successful troubleshooting.