Puritan Bennett 980

Network or Data Export Communication Failure

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

980

What This Guide Helps With

Troubleshooting failed network communication or data export caused by disconnected cables, inactive ports, configuration errors, removable media, or external system problems.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not interrupt ventilation or restart the Puritan Bennett 980 while it is actively supporting a patient solely to troubleshoot a communication problem.

Expected outcome: Patient ventilation remains uninterrupted while the communication problem is evaluated safely.

Continue Clinical Engineering troubleshooting only when the device can be evaluated without placing the patient at risk.

2. Identify the Exact Communication Failure

Determine which function is affected:

Record any displayed message, interface status, failed export notification, or downstream system alarm.

Expected outcome: The affected interface and failure pattern are clearly identified.

If the issue involves only one external destination, begin with that connection rather than assuming the ventilator has failed.

3. Confirm Normal Ventilator Operation

Verify that:

A broader startup or software problem may prevent communication interfaces from operating correctly.

Expected outcome: The problem is isolated to communication or data export rather than general ventilator operation.

If the ventilator shows other operational failures, remove it from service and address those failures first.

4. Inspect External Communication Connections

Inspect all applicable external connections for:

Reseat each connection only after confirming that doing so will not disrupt active patient care.

Expected outcome: All cables are correctly connected and free of visible damage.

If reseating a loose cable restores communication, verify stable data transfer and stop.

5. Verify the Correct Interface and Cable Type

Confirm that the connected cable and interface match the hospital integration design.

Check whether the system is intended to use:

Do not assume that a cable fitting physically means it is wired, configured, or supported correctly.

Expected outcome: The ventilator is connected through the intended port using the correct cable and approved interface equipment.

If the wrong cable, adapter, or port was used, correct the connection and retest.

6. Test the External Cable

Replace the communication cable with a known-good cable of the same approved type.

For Ethernet connections, confirm that:

For serial connections, confirm that:

Expected outcome: Communication returns with a known-good cable, identifying the original cable as the cause.

If the replacement cable does not restore communication, continue.

7. Check the External Network or Integration Device

Inspect the connected external equipment, such as:

Confirm that the external device has power and shows its expected status indicators.

Connect the PB980 to a known-good approved network or gateway port when facility policy permits. Alternatively, connect a known-good compatible device to the original port to determine whether the external infrastructure is functioning.

Expected outcome: The problem is isolated either to the ventilator or to the external network and integration path.

If another device also fails on the same connection, escalate the network or integration issue to the responsible IT or clinical systems team.

8. Verify Network Configuration

Coordinate with the facility’s IT or device-integration team to confirm the assigned network configuration.

Check, as applicable:

Do not change network parameters without authorization and documented configuration information.

Expected outcome: The ventilator’s communication settings match the approved hospital integration configuration.

If correcting an authorized configuration error restores communication, verify data receipt at the destination and stop.

9. Confirm Destination-System Status

Verify that the receiving system is available and actively accepting data.

Ask the responsible support team to confirm:

A ventilator may transmit correctly while the downstream system fails to display or store the information.

Expected outcome: The receiving system is confirmed operational or identified as the cause.

If the downstream system is responsible, document the finding and transfer the incident to the appropriate support group.

10. Evaluate USB Data Export Separately

When the reported problem involves USB export:

Do not connect an unknown or unapproved USB device to the ventilator.

Medtronic previously issued a PB980 software update addressing external USB-drive performance and its potential effect on graphical user interface operation. Software revision and applicable field actions should therefore be verified when USB behavior is abnormal.

Expected outcome: The ventilator recognizes the approved storage device and completes the export successfully.

If one USB device works and another does not, remove the incompatible or failed storage device from use.

11. Perform a Controlled Restart When Safe

If the communication interface remains unresponsive and the ventilator is not supporting a patient:

Do not repeatedly power-cycle the ventilator.

Expected outcome: A temporary software or interface-state problem clears and communication resumes.

If communication returns, perform repeated connection or export checks before returning the ventilator to service.

12. Check Software Revision and Service History

Review the ventilator’s:

The PB980 has had manufacturer software updates affecting external USB-drive behavior, so outdated software or incomplete field-action implementation should be considered during evaluation.

Do not install or alter ventilator software unless authorized, trained, and following current manufacturer service documentation.

Expected outcome: The software revision and field-action status are confirmed current or referred for authorized correction.

13. Verify Communication Before Return to Service

After correcting the problem:

Expected outcome: Communication remains reliable and the ventilator passes all required return-to-service checks.

If communication remains intermittent or cannot be verified, do not return the ventilator to service.

If the Problem Persists

If known-good cables, external ports, approved media, configuration settings, receiving systems, and a controlled restart have been checked, common external causes have been ruled out.

The failure may involve an internal communication interface, connector assembly, software fault, or electronic hardware problem.

The ventilator should be:

Do not open the ventilator or perform board-level troubleshooting without current service documentation, proper training, and authorization.

Knowing when external troubleshooting is complete and escalation is required is proper Clinical Engineering troubleshooting.

Clinical Use Tip

A communication failure may not stop ventilation, but it can interrupt charting, remote data visibility, alarm distribution, or clinical workflow. Never allow network availability to substitute for direct observation of the ventilator and patient.

Do not restart, disconnect, or reconfigure a PB980 while it is supporting a patient unless the patient has first been safely transferred according to facility procedure.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Puritan Bennett 980 was ventilating normally but no patient data appeared in the device-integration system."

Cause

What was observed during troubleshooting.

Example:
"Inspection found a damaged Ethernet cable locking tab that allowed the connector to intermittently disconnect from the ventilator."

Resolution

What action was taken.

Example:
"Replaced the Ethernet cable, verified stable network communication and data receipt at the integration system, completed functional checks, and returned the ventilator to service."

Helpful Details to Include (If Known)

Final Thought

Patient safety comes first even when the ventilator continues operating normally. A logical evaluation should separate the ventilator, cable, network, removable media, and receiving system before internal failure is suspected. Proper escalation prevents unnecessary disassembly, while complete CCR documentation preserves the technical findings and final equipment status.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide