Puritan Bennett 980

O₂ Supply Failure or Delivered FiO₂ Incorrect

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

980

What This Guide Helps With

Troubleshooting oxygen-supply alarms or incorrect delivered FiO₂ caused by gas sources, hoses, inlet connections, settings, accessories, or external supply problems.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a suspected oxygen-delivery failure while the Puritan Bennett 980 is actively supporting a patient.

Expected outcome: The patient is safely supported without depending on a ventilator with uncertain oxygen delivery.

Continue Clinical Engineering troubleshooting only after the PB980 has been removed from patient use.

2. Confirm the Exact Reported Problem

Review the alarm history and determine whether the problem involves:

Record the set oxygen percentage, displayed oxygen percentage, alarm wording, operating mode, and whether the condition is continuous or intermittent.

Expected outcome: The failure is clearly defined before parts or accessories are changed.

3. Verify Both Gas Sources Are Connected

Confirm that both medical-grade oxygen and compressed air are connected to the correct PB980 inlet fittings.

The PB980 mixes oxygen and air internally to produce the selected oxygen concentration. The manufacturer warns that relying on only one gas source can result in loss of ventilation or hypoxemia if that source fails.

Expected outcome: Both gas sources are securely connected to their correct inlets.

If reconnecting the hoses resolves the alarm and FiO₂ verifies correctly, complete functional testing and stop.

4. Check the Facility Oxygen Source

Test the oxygen outlet with an approved pressure gauge or known-good device according to facility policy.

Check for:

Do not assume the ventilator is defective until the facility oxygen source has been verified.

Expected outcome: The oxygen outlet provides stable pressure within the requirements listed for the installed PB980 configuration and hose system.

If the outlet is defective, remove it from clinical use and notify Facilities or the responsible medical-gas service.

5. Test the Ventilator at Another Approved Oxygen Outlet

Move the PB980 to a known-good medical-gas outlet when practical.

Expected outcome: The test separates a facility outlet problem from a ventilator or hose problem.

If the PB980 operates normally at another outlet, the original wall source is the likely cause. Stop ventilator troubleshooting and document the outlet issue.

6. Inspect the Oxygen Hose and Fittings

Disconnect the oxygen hose after safely isolating the gas supply and inspect it for:

Verify that the hose is rated and approved for medical oxygen service.

Expected outcome: The oxygen hose and fittings are intact, unobstructed, correctly configured, and leak-free.

7. Substitute a Known-Good Oxygen Hose

Replace the oxygen hose with a compatible, approved, known-good hose.

Reconnect the PB980 to a verified oxygen source and observe:

Some hose assemblies may create excessive restriction at lower supply pressures, reducing ventilator performance.

Expected outcome: Normal operation with the replacement hose identifies the original hose or fitting as defective.

If the hose resolves the problem, remove the failed hose from service and stop.

8. Inspect the PB980 Oxygen Inlet Externally

Examine the ventilator oxygen inlet without opening the ventilator.

Check for:

Do not insert tools into the gas inlet or attempt internal valve repair during basic troubleshooting.

Expected outcome: The inlet is secure, clean, undamaged, and retains the oxygen connection correctly.

Visible inlet damage requires removal from service and repair evaluation.

9. Verify the Set Oxygen Percentage

Confirm that the selected O₂ percentage matches the intended test condition.

Check for:

The PB980’s mix module blends air and oxygen and uses an oxygen sensor to monitor the resulting mixture against the operator-selected O₂ setting.

Expected outcome: The selected oxygen percentage is appropriate and the monitored value is being interpreted correctly.

10. Inspect the Patient Circuit and Test Setup

Verify that the ventilator is connected to the correct circuit and test lung for the selected patient category.

Check for:

Circuit or test-setup problems may create unstable readings or misleading analyzer results.

Expected outcome: The patient circuit and test setup are complete, sealed, and appropriate for testing.

11. Complete the Required Pre-Use or System Check

Run the PB980’s approved pre-use checkout or short self-test using the correct circuit and test equipment.

Do not bypass failed test steps.

Observe whether the ventilator identifies:

Expected outcome: The ventilator completes the required checkout without gas-delivery or oxygen-monitoring failures.

If the checkout fails repeatedly after external causes are corrected, remove the ventilator from service.

12. Compare Delivered FiO₂ With an Independent Analyzer

Connect a calibrated oxygen analyzer according to facility test procedures.

Test several representative oxygen settings, such as:

Allow readings to stabilize before recording results. Use a controlled test lung and verify that both gas supplies remain stable during the test.

Compare:

Expected outcome: The independent analyzer and ventilator reading remain within the applicable manufacturer performance specification and facility test tolerance.

A large or unstable discrepancy suggests an oxygen-sensor, gas-mixing, inlet-regulation, or internal pneumatic problem.

13. Evaluate Whether the Problem Follows the Gas Source

When safe and permitted, repeat the test using:

Change only one item at a time so the cause can be identified.

Expected outcome: The technician determines whether the failure follows the outlet, hose, gas source, or ventilator.

14. Review Alarm and Diagnostic Information

Record all active alarms, diagnostic codes, timestamps, and repeated failure patterns.

Determine whether the PB980 reports:

Do not clear diagnostic information before it has been documented.

Expected outcome: Sufficient information is preserved for bench troubleshooting, service support, or repair escalation.

15. Perform Final Functional Verification

If an external cause was corrected:

Do not return the PB980 to service based only on the disappearance of an alarm.

Expected outcome: The ventilator passes all required functional and oxygen-delivery testing before release.

If the Problem Persists

If verified gas sources, wall outlets, hoses, connections, settings, circuit components, and test equipment have been ruled out, the problem is likely internal.

Possible internal causes may include the oxygen sensor, inlet regulation components, gas flow sensing, gas-mixing components, pneumatic valves, wiring, or control electronics.

The ventilator should be:

Do not continue operating a PB980 that cannot reliably deliver or verify the selected FiO₂. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot uncertain oxygen delivery while the PB980 is connected to an active patient. Transfer the patient first and independently verify oxygenation on the replacement equipment.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported repeated oxygen-supply alarms and delivered FiO₂ approximately 10% below the selected value."

Cause

What was observed during troubleshooting.

Example:
"The oxygen wall hose was internally restricted; the PB980 operated normally with a verified outlet and known-good replacement hose."

Resolution

What action was taken.

Example:
"Removed the defective oxygen hose from service, installed an approved replacement, completed system checkout, and verified multiple FiO₂ settings with a calibrated oxygen analyzer."

Helpful Details to Include (If Known)

Final Thought

Oxygen-delivery complaints require immediate attention because inaccurate FiO₂ can directly affect patient safety. Verify external gas sources, hoses, settings, and test equipment logically before suspecting an internal failure. Escalate when oxygen delivery cannot be confidently verified, and document the complete troubleshooting path.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide