Puritan Bennett 980 Ventilator

O₂ Sensor Errors / Low or Unstable FiO₂

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

980

What This Guide Helps With

This guide addresses situations where the Puritan Bennett 980 ventilator reports O₂ sensor errors, low FiO₂ readings, or unstable oxygen delivery.

The focus is on logical, external troubleshooting first—such as oxygen supply, wall source pressure, blenders, filters, and calibration—before assuming internal sensor or board failure.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify Oxygen Wall Source

3. Inspect the High-Pressure O₂ Hose

4. Confirm Air Supply

5. Check for Active O₂ Sensor Calibration Alerts

6. Inspect the O₂ Sensor Module

7. Check for Excessive Moisture in the System

8. Verify Alarm Behavior

If the Problem Persists

If oxygen source pressure, hoses, calibration, moisture, and sensor replacement have been ruled out, the issue is likely internal (blender module, internal sensor circuitry, or control board).

Knowing when to stop is proper troubleshooting. Avoid unnecessary internal disassembly in the clinical environment.

Clinical Use Tip

Never troubleshoot oxygen delivery issues on an actively ventilated patient unless clinically stable and monitored.

If FiO₂ accuracy is in question, move the patient to another verified ventilator first. Oxygen delivery is life-sustaining—verification is critical.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator alarming ‘O₂ Sensor Error’ with low FiO₂ reading at 40% despite 60% setting.”

Cause

What was observed during troubleshooting.

Example:
“Found O₂ pipeline hose leaking at DISS fitting. FiO₂ unstable due to fluctuating oxygen pressure.”

Resolution

What action was taken.

Example:
“Replaced high-pressure O₂ hose, verified 50 psi pipeline pressure, recalibrated O₂ sensor. FiO₂ stabilized and passed operational check.”

“Replaced O₂ sensor, completed successful calibration, verified FiO₂ with external analyzer. Unit returned to service.”

Helpful Details to Include

Final Thought

Oxygen delivery problems must always be approached methodically. Start with gas supply and simple connections before assuming internal failure. Protect the patient first, verify performance second, and escalate appropriately when external causes are ruled out. Clear documentation protects both patients and the department.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide