Puritan Bennett 980

Ventilator Inoperative or Safety Valve Open Condition

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

980

What This Guide Helps With

Troubleshooting Ventilator Inoperative or Safety Valve Open conditions caused by power, gas-supply, circuit, startup, accessory, or self-test failures.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a Ventilator Inoperative or Safety Valve Open condition while the Puritan Bennett 980 is supporting a patient.

A Safety Valve Open condition may permit spontaneous breathing through the ventilator, but normal assisted ventilation may not be available. Do not rely on this condition for continued patient support.

Expected outcome: The patient is safely supported without relying on the affected ventilator.

Continue Clinical Engineering troubleshooting only after the PB980 has been removed from patient use.

2. Record the Exact Alarm and Operating State

Observe and document:

Do not repeatedly reset the ventilator before recording the available fault information.

Expected outcome: The failure condition and supporting alarm information are clearly documented.

3. Verify AC Power

If the condition clears after restoring reliable AC power, complete operational testing before returning the ventilator to service.

Expected outcome: Stable AC power is available and recognized by the ventilator.

4. Verify Both Gas Supplies

The PB980 normally requires properly connected medical air and oxygen supplies.

Loss of both compressed-gas sources can place the PB980 into a Safety Valve Open condition. Reconnecting an adequate gas source may allow recovery.

If restoring the gas supply clears the condition, continue with self-test and functional verification before clinical use.

Expected outcome: Both air and oxygen supplies are present, correctly connected, and free of obvious restrictions.

5. Inspect the Patient Circuit and External Accessories

With the ventilator removed from patient use:

A severe restriction, incorrect circuit assembly, or improperly installed expiratory component may prevent startup or normal ventilation.

Expected outcome: A correctly assembled, unobstructed test circuit is securely connected.

6. Perform a Controlled Power Restart

Only after the patient has been transferred and external power, gas, and circuit conditions have been checked:

Do not repeatedly power-cycle a ventilator that immediately returns to Ventilator Inoperative.

If the ventilator starts normally, continue with self-test and complete functional verification.

Expected outcome: The ventilator completes startup without returning to the critical fault condition.

7. Run the Short Self-Test

Using the approved test circuit, stopper, and accessories:

The PB980 operator documentation identifies the Short Self-Test as part of patient-circuit setup and verification.

If the test passes and the fault does not recur, complete an extended operational check before returning the device to service.

Expected outcome: The ventilator passes the Short Self-Test without Ventilator Inoperative or Safety Valve Open indications.

8. Check for Obvious Environmental or Physical Problems

Expected outcome: No external physical or environmental condition explains the failure.

9. Review Alarm and Diagnostic History

When accessible:

Do not enter restricted service functions or clear diagnostic history unless authorized by facility policy and manufacturer training.

Expected outcome: Relevant fault information is preserved for repair evaluation.

10. Complete Final Functional Verification

If the condition was corrected by an external action:

Do not return the PB980 to clinical service based only on the disappearance of the alarm.

Expected outcome: The ventilator operates reliably, passes required testing, and is safe for clinical use.

If the Problem Persists

If the Puritan Bennett 980 continues to display Ventilator Inoperative or Safety Valve Open after power, gas supplies, hoses, circuit components, accessories, startup conditions, and approved self-testing have been checked, common external causes have been ruled out.

The condition may involve an internal gas-delivery, pressure-control, power, communication, processor, valve, sensor, or safety-system failure.

The ventilator should be:

Do not open internal pneumatic or electronic assemblies without the required training, service documentation, tools, and test equipment. Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never investigate a Ventilator Inoperative or Safety Valve Open condition while the PB980 remains responsible for patient ventilation. Transfer the patient first and verify the replacement ventilator before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Puritan Bennett 980 entered Ventilator Inoperative with Safety Valve Open and stopped providing normal assisted ventilation."

Cause

What was observed during troubleshooting.

Example:
"Both compressed-gas hoses were connected, but the medical-air hose fitting was not fully seated and the ventilator could not maintain the required gas supply."

Resolution

What action was taken.

Example:
"Reseated and verified both gas connections, installed a known-good test circuit, completed the Short Self-Test and functional verification, and returned the ventilator to service after all tests passed."

Helpful Details to Include (If Known)

Final Thought

A Ventilator Inoperative or Safety Valve Open condition must be treated as a loss of dependable ventilatory support. Protect the patient first, check external power, gas, circuit, and setup conditions logically, and escalate persistent failures for qualified repair. Accurate CCR documentation preserves critical information and supports an efficient, defensible repair process.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide