Puritan Bennett 980

Will Not Power On or Fails During Startup

On this page

Asset Type

Ventilator

Manufacturer

Puritan Bennett

Model

980

What This Guide Helps With

Troubleshooting no-power conditions, blank displays, interrupted startup, repeated rebooting, startup alarms, or failure to reach the normal ventilator setup screen.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Do not troubleshoot a Puritan Bennett 980 that will not start or complete startup while it is connected to a patient.

Expected outcome: The patient is safely supported without relying on a ventilator that cannot complete its startup sequence.

Continue Clinical Engineering troubleshooting only after the affected PB980 has been removed from patient use.

2. Confirm the Exact Failure

Attempt to identify and document whether the ventilator:

Observe both the graphical user interface and the status display on the breath delivery unit. The PB980 incorporates a separate status display that can provide additional indication of ventilator operation.

Expected outcome: The failure is narrowed to total power loss, display-related behavior, interrupted startup, or a reported system fault.

3. Inspect for Immediate Safety Hazards

Before reconnecting power, inspect the ventilator for:

Do not energize the unit when liquid intrusion, burning odor, smoke, severe physical damage, or electrical damage is present.

Expected outcome: No immediate electrical, fire, or mechanical hazard is found.

If a hazard is present, stop troubleshooting, remove the ventilator from service, and send it for controlled bench evaluation.

4. Verify the AC Power Source

Disconnect the ventilator and test the wall receptacle with an approved outlet tester or another known-good device.

Confirm that:

Avoid relying solely on the report that “the outlet works.”

Expected outcome: The AC source is verified as functional and appropriate.

If the outlet fails testing, move the ventilator to a verified receptacle. If the ventilator then starts normally, stop and document the facility power problem.

5. Inspect and Reseat the Power Cord

Disconnect AC power and inspect the full length of the power cord.

Expected outcome: The ventilator has a secure, undamaged connection to verified AC power.

If reseating or replacing the cord restores normal startup, complete the required operational and safety checks before returning the unit to service.

6. Disconnect Nonessential External Accessories

Power the ventilator off and disconnect nonessential external devices, including:

Leave only the approved power connection and equipment necessary for a controlled startup evaluation.

Expected outcome: An external accessory or damaged peripheral connection is ruled out as a cause of startup interruption.

If the ventilator starts normally, reconnect accessories individually to identify the device or cable associated with the failure. Stop using the suspect accessory.

7. Check Battery and AC-Power Indications

Connect the ventilator to verified AC power and observe all available indicators.

Determine whether:

The PB980 supports battery operation, but battery availability must not be treated as proof that the AC input or internal charging system is functioning correctly. Medtronic describes the PB980 as using hot-swappable batteries, depending on configuration.

Expected outcome: Clinical Engineering determines whether incoming AC power, stored battery power, or both are being recognized.

Do not repeatedly cycle power from a severely depleted or questionable battery.

8. Allow Adequate AC Charging Time When the Battery Is Depleted

If the ventilator has been stored, unplugged, or unused for an extended period:

Do not assume a fully depleted battery is the only problem if the ventilator also fails to respond while connected to verified AC power.

Expected outcome: A deeply discharged battery condition is either corrected or ruled out.

If the ventilator starts after charging, evaluate battery condition and runtime before returning the unit to clinical use.

9. Perform One Controlled Power Cycle

If no electrical hazard is present:

Record:

Expected outcome: A temporary startup state clears, or the failure becomes repeatable and documentable.

If the ventilator completes startup normally, continue with all required pre-use testing before considering it operational.

10. Verify the Graphical User Interface Connection Externally

Inspect the graphical user interface, mounting arm, and externally accessible cable path without removing covers.

Check for:

Do not open the graphical user interface or breath delivery unit during basic troubleshooting.

Expected outcome: Obvious external display-connection damage is identified or ruled out.

A working breath delivery unit status display with an inactive graphical interface may indicate a display, cable, communication, or internal power problem requiring bench evaluation.

11. Record Startup Messages and Diagnostic Codes

Photograph or transcribe any startup message exactly as displayed.

Document:

Do not clear logs, perform unauthorized software changes, or repeatedly reboot the ventilator in an attempt to bypass a persistent fault.

Expected outcome: Repair personnel receive precise information that can be correlated with manufacturer service documentation.

12. Complete Required Testing After a Successful Startup

A ventilator that eventually starts after an unexplained failure should not immediately be returned to patient use.

Complete the facility-approved inspection and applicable manufacturer testing, including:

Medtronic directs users to the PB980 operator’s manual for complete operating instructions, warnings, and pre-use requirements.

Expected outcome: The ventilator consistently starts, passes required testing, and performs normally before being released.

If any test fails or startup remains intermittent, remove the ventilator from service.

If the Problem Persists

If verified AC power, the power cord, batteries, accessories, external connections, and a controlled power cycle do not restore reliable startup, common external causes have been ruled out.

The failure may involve an internal power supply, battery interface, graphical user interface, breath delivery unit, internal communication path, cooling system, software, or another protected component.

The ventilator should be:

Do not return a ventilator to clinical use when startup is intermittent, a technical fault remains active, or the cause of an unexpected shutdown has not been established.

Knowing when to stop and escalate is proper troubleshooting.

Clinical Use Tip

Never troubleshoot a PB980 startup or power failure while it is the only ventilator supporting an active patient. Move the patient to verified equipment first.

A ventilator that starts after several attempts is not necessarily reliable. Intermittent startup behavior should be treated as a potential recurrence risk until the cause is identified and required testing is passed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Respiratory therapy reported that the Puritan Bennett 980 remained blank and would not complete startup when needed for patient use."

Cause

What was observed during troubleshooting.

Example:
"Verified wall power and power cord were functional, but the ventilator repeatedly stopped during startup with the same technical fault displayed."

Resolution

What action was taken.

Example:
"Removed the PB980 from service, labeled it Out of Service, documented the startup fault, and routed it for authorized bench evaluation."

Helpful Details to Include (If Known)

Final Thought

Patient safety comes before restoring power. A logical evaluation of the outlet, power cord, batteries, accessories, indicators, and startup behavior prevents unnecessary disassembly and supports appropriate escalation. Clear CCR documentation gives repair personnel the information needed to reproduce the failure and verify a safe resolution.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide