ResMed Astral Ventilator

Circuit Leak or Disconnection Errors

On this page

Asset Type

Ventilator

Manufacturer

ResMed

Model

Astral

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting circuit leak or disconnection errors on a ResMed Astral ventilator. Symptoms may include alarms indicating “circuit leak,” “disconnection,” or unexpected patient pressure/volume changes. The focus is on external, verifiable checks first, ensuring patient safety and avoiding unnecessary internal interventions.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Check Circuit Connections

Inspect Circuit Components

Check the Patient Interface (Mask/ET Tube)

Verify Ventilator Settings

Check Humidifiers or Water Traps

If the Problem Persists

All external connections, tubing, and patient interfaces have been verified. Internal circuit sensors or ventilator modules may be malfunctioning.

Knowing when to escalate avoids unnecessary patient risk and prevents further device damage.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator alarm indicates circuit leak or disconnection; patient pressure readings unstable.”

Cause

What was observed during troubleshooting.

Example:
“Loose tubing connection detected at patient Y-piece; mask cushion slightly torn.”

Resolution

What action was taken.

Example:
“Reconnected all circuit tubing; replaced mask cushion; verified ventilator circuit test passed; monitored patient for stability.”

Helpful Details to Include

Final Thought

Patient safety is the priority—always verify proper ventilation before troubleshooting. Systematic external checks reduce unnecessary repair escalation. Thorough documentation ensures continuity of care and supports proper clinical engineering workflow.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide