ResMed Astral Ventilator

Flow or Tidal Volume Discrepancies

On this page

Asset Type

Ventilator

Manufacturer

ResMed

Model

Astral

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting flow or tidal volume discrepancies on a ResMed Astral ventilator. Symptoms may include delivered tidal volumes being higher or lower than set, inconsistent flow patterns, unexpected alarms related to volume, or patient-ventilator asynchrony. The guide focuses on external and verifiable checks before assuming internal malfunction.

Step-by-Step Troubleshooting

Verify Patient Circuit Connection

Inspect Circuit for Leaks

Check Exhalation Port and Filters

Verify Ventilator Settings vs Patient Requirements

Inspect for Water or Condensation in Circuit

Check for Kinks or Obstructions in Tubing

Examine Ventilator Battery or Power Supply

Replace Disposable Components

If the Problem Persists

After completing all external checks, if discrepancies continue, the issue is likely internal (sensor malfunction, flow sensor miscalibration, or board-level error).

Clinical Use Tip

Never troubleshoot flow or volume issues while the patient is connected to the ventilator if alternate ventilation is not immediately available.

Move the patient to a backup ventilator before testing.

Confirm alarms are functioning properly before returning the device to clinical use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Clinician reports tidal volume delivered is inconsistent with set values; low and high tidal volume alarms observed.”

Cause

What was observed during troubleshooting.

Example:
“External inspection revealed circuit and filters intact, no visible leaks; flow discrepancies persist.”

Resolution

What action was taken.

Example:
“Ventilator removed from service, labeled Out of Service, and sent for manufacturer repair. Patient transitioned to backup device.”

Helpful Details to Include

Final Thought

Accurate tidal volume delivery is critical for patient safety. By following a logical, stepwise approach—checking connections, leaks, and circuit components—Clinical Engineering can rule out common external causes efficiently. Escalation to repair should be done promptly if discrepancies persist, ensuring patient safety and equipment reliability. Proper CCR documentation supports accountability and continuous quality improvement.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide