ResMed Lumis Ventilator

Power or Battery Failure

On this page

Asset Type

Ventilator

Manufacturer

ResMed

Model

Lumis

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting power or battery failures on a ResMed Lumis ventilator. Symptoms may include the device not turning on, failure to charge, rapid battery drain, or unexpected shutdowns. The focus is on external checks and verifiable causes before escalating to internal repair.

Step-by-Step Troubleshooting

Verify AC Power Source

Check the Ventilator’s Power Switch and Indicators

Inspect the Battery Module

Test with a Known Good Battery

Check Charging Behavior

Power Cycle the Device

If the Problem Persists

All external power sources, battery modules, and cabling have been ruled out.

The device may have an internal power supply or board failure.

Action:

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Ventilator will not power on; battery not charging.”

Cause

What was observed during troubleshooting.

Example:
Battery was found loose and corroded; AC power confirmed functional.

Resolution

What action was taken.

Example:
Cleaned battery contacts, reseated battery, and verified device powers on and charges. If unresolved, device sent for manufacturer repair.

Helpful Details to Include

Final Thought

Logical, stepwise troubleshooting ensures patient safety while avoiding unnecessary internal repairs. Proper documentation and timely escalation protect both the patient and the device. Always confirm the ventilator is functioning fully before returning to service.

That is successful troubleshooting.