Roche cobas pro

Clot, Bubble, or Sample-Volume Error Repeats

On this page

Asset Type

Clinical Chemistry Analyzer

Manufacturer

Roche

Model

cobas pro

What This Guide Helps With

Troubleshoots repeated sample-quality or volume errors caused by specimen condition, tube selection, placement, aspiration access, or recurring analyzer sampling problems.

Step-by-Step Troubleshooting

1. Protect Patient Results and Identify the Affected Specimen

Do not repeatedly aspirate a patient specimen when the analyzer is reporting clot, bubble, or insufficient-volume conditions.

Determine:

Follow laboratory procedure for questionable or partially processed samples.

Expected outcome: The affected specimen is identified and unsafe repeat processing is avoided.

2. Inspect the Specimen Externally

Have laboratory personnel evaluate the specimen for:

Clinical Engineering should not independently alter or reprocess patient specimens outside laboratory procedure.

Expected outcome: Obvious specimen-related causes are identified or ruled out.

3. Verify Tube Type and Placement

Confirm with laboratory staff that:

Expected outcome: The sample container is positioned correctly and can be accessed normally.

4. Compare With Another Approved Sample

Determine whether a different verified specimen or laboratory-approved test sample runs normally in the same rack or loading path.

Expected outcome: Normal processing of another sample suggests the original specimen or container is the primary cause.

5. Inspect the Rack or Carrier

Check for:

Replace a suspect rack with a known-good compatible rack if appropriate.

Expected outcome: The carrier maintains the sample in the correct aspiration position.

6. Check for Repeated Errors Across Multiple Samples

If multiple unrelated specimens show similar errors, verify:

A broader pattern makes a single specimen problem less likely.

Expected outcome: The issue is correctly classified as specimen-specific or analyzer-wide.

7. Inspect Accessible Sampling Areas

With the analyzer safely stopped according to normal procedures, inspect visible sampling areas for:

Use only approved cleaning procedures.

Expected outcome: Accessible sampling areas are clean and visibly intact.

8. Verify Corrected Operation

After the laboratory has corrected or replaced the specimen as appropriate:

Expected outcome: The replacement or corrected sample processes normally. Troubleshooting can stop when consistent aspiration is demonstrated.

9. Escalate Analyzer-Wide Repeated Errors

If properly prepared samples with sufficient volume repeatedly trigger the same type of error, stop external troubleshooting.

Do not adjust internal pressure, sensing, aspiration, or probe calibration systems without authorized procedures.

Expected outcome: A potential analyzer sampling or sensing problem is escalated.

If the Problem Persists

Common specimen, container, carrier, volume, and accessible contamination causes have been ruled out. The remaining problem may involve aspiration sensing, pipetting, pressure detection, fluidics, probe positioning, or service-level calibration.

The affected analyzer or module should be:

Required functional checks and laboratory QC must be completed before return to patient testing.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

When the error follows one specimen, preserve specimen integrity and involve laboratory staff before repeatedly rerunning or manipulating the sample.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported repeated low-volume errors on one specimen despite multiple loading attempts."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the tube contained insufficient accessible sample volume for reliable aspiration."

Resolution

What action was taken.

Example:
"Laboratory staff provided an appropriate replacement specimen, Clinical Engineering verified normal sample handling, and the analyzer completed processing without repeat aspiration errors."

Helpful Details to Include (If Known)

Final Thought

Repeated sample-quality errors should be separated from analyzer problems by checking the specimen, tube, carrier, and aspiration path in order. Preserve sample integrity, verify successful processing, and escalate broader repeated failures.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide