Roche cobas pro

Probe, Pipettor, or Aspiration Error Occurs

On this page

Asset Type

Clinical Chemistry Analyzer

Manufacturer

Roche

Model

cobas pro

What This Guide Helps With

Troubleshoots probe, pipetting, and aspiration errors caused by sample condition, positioning, consumables, obstruction, contamination, or fluid-supply problems.

Step-by-Step Troubleshooting

1. Protect Patient Testing and Stop Repeated Aspiration Attempts

Do not continue patient testing when reliable sample or reagent aspiration cannot be assured.

Identify:

If there is evidence of a probe collision, bent component, spill, or fluid leak, stop operation and remove the affected module from service.

Expected outcome: Testing is safely paused and the exact aspiration-related condition is identified.

2. Inspect the Sample or Reagent Container

Check for:

Expected outcome: The fluid being aspirated is physically suitable for normal analyzer access.

3. Verify Rack and Container Seating

Confirm:

Use a known-good carrier when appropriate.

Expected outcome: The probe has an unobstructed path to a correctly positioned container.

4. Inspect Accessible Probe Areas

With the analyzer safely stopped according to normal procedures, visually inspect accessible probe areas for:

Do not manually straighten, realign, or disassemble a damaged probe unless specifically authorized and trained under approved service procedures.

Expected outcome: No obvious external damage or contamination is present. Visible damage requires removal from service and escalation.

5. Check Wash and System Fluid Conditions

Verify externally accessible:

Correct only approved external conditions.

Expected outcome: Required external fluid supplies and waste pathways are available.

6. Determine Whether the Problem Follows One Sample

When appropriate, compare the affected specimen with a verified test sample or laboratory-approved known-good specimen presentation.

If only one specimen repeatedly fails, have laboratory personnel evaluate:

Expected outcome: Sample-specific conditions are separated from analyzer-wide aspiration problems.

7. Verify Operation After Correcting the External Cause

After correcting a legitimate external problem, run the analyzer's normal approved workflow.

Observe for:

Expected outcome: Aspiration completes without repeat errors. If normal operation is restored, proceed to final verification.

8. Perform Final Functional Verification

Confirm:

Expected outcome: Reliable pipetting and aspiration are demonstrated and laboratory acceptance requirements are met. Troubleshooting is complete.

9. Escalate Repeated or Mechanical Errors

Stop troubleshooting if:

Expected outcome: Potential internal mechanical or fluidic problems are escalated before further damage or unreliable testing occurs.

If the Problem Persists

Common sample, container, rack, fluid-supply, and accessible contamination causes have been ruled out. The remaining condition may involve internal probe alignment, pumps, valves, pressure sensing, motion systems, tubing, control electronics, or service-level calibration.

The analyzer or affected module should be:

Required functional, calibration, and QC verification must be completed before patient testing resumes.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

A probe collision or repeated aspiration fault can threaten both analyzer hardware and result integrity; do not keep rerunning specimens to see if it clears.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported repeated aspiration errors when the cobas pro attempted to process specimens from one rack."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found one sample tube was not fully seated and was positioned abnormally in the rack."

Resolution

What action was taken.

Example:
"Corrected the tube position, inspected the accessible probe area for damage, verified normal aspiration, and confirmed acceptable laboratory QC before return to service."

Helpful Details to Include (If Known)

Final Thought

Aspiration faults require protection of both patient results and the analyzer's sampling mechanism. Check samples, positioning, supplies, and visible conditions first, then stop before internal mechanical repair when reliable aspiration cannot be restored.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide