Roche cobas pro

Result Is Delayed, Missing, or Inconsistent

On this page

Asset Type

Clinical Chemistry Analyzer

Manufacturer

Roche

Model

cobas pro

What This Guide Helps With

Troubleshoots delayed, missing, or inconsistent results caused by sample status, analyzer holds, QC, calibration, communication, workflow, or repeat-processing conditions.

Step-by-Step Troubleshooting

1. Protect Patient Care and Define the Result Problem

Determine whether the result is:

Notify laboratory personnel of any potentially significant reporting delay and follow established critical-result or downtime procedures.

Do not release or reconstruct a result based on assumption.

Expected outcome: The missing or inconsistent result is traced to a specific specimen, assay, and workflow stage.

2. Verify Positive Specimen Identification

Confirm:

Resolve identification discrepancies before further troubleshooting.

Expected outcome: The specimen being investigated is unquestionably matched to the correct order and patient record.

3. Check Analyzer Sample Status

Review normal analyzer workflow information to determine whether the specimen is:

Record any message rather than repeatedly resubmitting the sample.

Expected outcome: The analyzer's current handling state is known.

4. Check Reagent, Calibration, and QC Status

Determine whether the assay is delayed or withheld because of:

Clinical Engineering should not override laboratory controls designed to prevent release of potentially invalid results.

Expected outcome: Any legitimate analytical hold is identified.

5. Inspect the Specimen and Sample Handling

Check with laboratory staff for:

Review whether the analyzer logged a sampling-related fault.

Expected outcome: Sample-specific causes of delay or inconsistency are identified or ruled out.

6. Determine Whether the Result Exists Locally

If the result is missing from the LIS, verify whether it is visible and finalized at the analyzer.

This separates:

Do not manually recreate or transmit a result unless authorized by laboratory procedure.

Expected outcome: The problem is localized to analysis, analyzer data handling, or downstream communication.

7. Check LIS and Network Communication

If the result is complete on the analyzer but absent downstream:

Expected outcome: The communication pathway is confirmed or the point of failure is identified.

8. Evaluate Inconsistent Results Carefully

For inconsistent results, have laboratory staff determine whether repeat testing is clinically and procedurally appropriate.

Clinical Engineering should investigate associated:

Do not conclude that the first or second result is correct based solely on which appears more plausible.

Expected outcome: Analyzer-related conditions that could affect consistency are identified without independently interpreting clinical validity.

9. Verify Complete Recovery

After correcting the identified cause:

Expected outcome: Results are produced, associated, and transmitted consistently. Troubleshooting is complete.

10. Escalate Unresolved or Unreliable Results

Stop external troubleshooting if:

Expected outcome: Potential result-integrity problems are prevented from affecting patient reporting and are escalated.

If the Problem Persists

Common specimen, reagent, calibration, QC, workflow, and external communication causes have been ruled out. Remaining possibilities include internal analytical systems, pipetting, measurement hardware, software, data handling, database synchronization, middleware, or service-level configuration.

The affected analyzer or assay should be:

Complete required calibration, QC, functional, and end-to-end result-transmission verification before return to patient testing.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Always determine whether a “missing result” is missing from the analyzer, merely held, or only missing from the LIS before troubleshooting the wrong system.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported completed cobas pro results were visible on the analyzer but were not appearing in the LIS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the analyzer's external network connection loose, interrupting result transmission while local analysis continued normally."

Resolution

What action was taken.

Example:
"Secured the network connection, verified interface restoration, and confirmed an approved test result transmitted successfully to the LIS before normal reporting resumed."

Helpful Details to Include (If Known)

Final Thought

Result problems require tracing the complete path from specimen identification through analysis and final transmission. Verify each stage before assuming internal failure, protect result integrity throughout, and escalate unresolved discrepancies with clear documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide