Roche cobas pro

Wash, Waste, or Fluidics System Fault Occurs

On this page

Asset Type

Clinical Chemistry Analyzer

Manufacturer

Roche

Model

cobas pro

What This Guide Helps With

Troubleshoots wash, waste, and fluidics faults caused by supply depletion, full waste containers, tubing issues, leaks, blocked drainage, or external utilities.

Step-by-Step Troubleshooting

1. Protect Patient Testing and Assess for Leakage

Pause affected testing immediately when fluid handling is unreliable.

Inspect around the analyzer for:

If active leakage creates an electrical, chemical, slip, or contamination hazard, isolate the analyzer and remove it from service.

Expected outcome: Immediate hazards are controlled and affected testing is redirected.

2. Confirm the Exact Fluidics Fault

Record:

Expected outcome: The affected external fluid function is identified before components are disturbed.

3. Check Accessible Supply Containers

Verify applicable external supplies are:

Confirm caps, pickup connections, and externally accessible tubing are positioned normally.

Expected outcome: Required fluids are available to the analyzer. If correcting an empty or poorly seated supply resolves the fault, proceed to verification.

4. Check Waste Containers and Drainage

Inspect accessible waste handling for:

Use proper PPE and follow laboratory biohazard procedures.

Expected outcome: Waste can leave the analyzer normally without leakage or backup.

5. Inspect Accessible Tubing and Connections

Check only externally accessible fluid lines for:

Do not open protected internal fluidics assemblies.

Expected outcome: Accessible tubing has a clear physical path and secure connections.

6. Check Facility Utilities When Applicable

If the analyzer depends on facility-provided water, drain, power, or other infrastructure, verify that the corresponding service is available.

Consider recent:

Expected outcome: Required facility utilities are available and functioning.

7. Use the Normal Approved Recovery Sequence

After correcting the external cause, allow the analyzer to perform only its normal approved prime, wash, startup, or recovery process.

Watch for:

Expected outcome: The analyzer completes its normal fluidics recovery without alarms or leaks.

8. Perform Final Verification

Confirm:

Expected outcome: Fluid handling is stable and required laboratory verification is successful. Troubleshooting is complete.

9. Escalate Persistent Fluidics Faults

Stop external troubleshooting if:

Expected outcome: The analyzer remains out of clinical service until qualified repair is completed.

If the Problem Persists

External supplies, waste handling, accessible tubing, drainage, and facility utilities have been checked. Remaining causes may involve internal pumps, valves, tubing, pressure or level sensors, fluid routing, wash mechanisms, or control systems.

The analyzer should be:

After repair, verify leak-free operation, normal fluidics functions, and required calibration/QC before return to service.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Treat any uncontrolled analyzer fluid leak as both a clinical-laboratory and equipment-safety issue until the source and fluid type are understood.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported the cobas pro stopped testing and displayed a waste-related fault."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external waste container full and not allowing normal waste handling."

Resolution

What action was taken.

Example:
"Laboratory staff emptied and correctly reseated the waste container using required PPE, Clinical Engineering verified normal fluidics operation with no leakage, and laboratory QC was acceptable."

Helpful Details to Include (If Known)

Final Thought

Fluidics troubleshooting begins with safety, supplies, waste, tubing, and facility utilities. Verify normal leak-free operation before assuming internal failure, and escalate promptly when the fault extends beyond accessible external systems.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide