Roche Urisys 1800

Test Strip Feed Or Positioning Error

On this page

Asset Type

Urine Analyzer

Manufacturer

Roche

Model

Urisys 1800

What This Guide Helps With

Troubleshooting test strip feed, strip transport, alignment, and positioning errors before removing the Urisys 1800 from service.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Stop Testing and Protect Result Integrity

Verify the Reported Error

Inspect the Test Strip Loading Area

Confirm Correct Strip Type

Check Strip Orientation and Placement

Check for Wet or Contaminated Strip Handling

Inspect the Strip Transport Area for Residue

Check Waste or Strip Collection Area

Restart the Analyzer if Appropriate

Run a Non-Patient Test or QC Check

Look for a Pattern

Check for Abnormal Mechanical Behavior

Do Not Force the Feed Mechanism

If the Problem Persists

If the Urisys 1800 continues to show test strip feed or positioning errors after strip type, strip handling, loading, waste area, cleaning, obstruction, and restart checks have been completed, common external causes have been ruled out.

The issue is likely related to an internal transport, sensor, motor, alignment, or control problem. The device should be removed from service, labeled Out of Service, and sent for repair or bench evaluation.

Knowing when to stop is proper troubleshooting. Continued testing on a device with unresolved strip positioning errors can risk delayed, incomplete, or unreliable urine results.

Clinical Use Tip

Do not troubleshoot the strip feed path while patient specimens are actively being processed. If urine testing is clinically needed, move testing to another validated analyzer or follow the lab’s approved downtime process before continuing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported the Roche Urisys 1800 displayed a test strip positioning error and would not complete urine strip analysis."

Cause

What was observed during troubleshooting.

Example:
"Inspection found dried urine residue and a partially obstructed used-strip exit area causing inconsistent strip advancement."

Resolution

What action was taken.

Example:
"Cleaned accessible strip path surfaces, cleared the used-strip area, verified correct strip loading, and confirmed normal operation with QC before returning the analyzer to service."

Helpful Details to Include (If Known)

Final Thought

For strip feed or positioning errors, start with patient safety and result integrity, then work through the visible strip path, reagent strips, loading technique, waste area, and cleaning condition. If the fault continues after external causes are ruled out, escalation for repair is the correct Clinical Engineering decision. Clear CCR documentation helps show what was checked, what was found, and why the device was returned to service or removed.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide