Samsung Healthcare GM85 Fit

Calibration or Quality-Control Test Fails

On this page

Asset Type

Mobile X-Ray System

Manufacturer

Samsung Healthcare

Model

GM85 Fit

What This Guide Helps With

Addresses failed calibration or QC caused by setup, detector condition, positioning, environment, power, test equipment, or persistent system faults.

Step-by-Step Troubleshooting

1. Protect Clinical Use

If the failed QC test affects confidence in image quality, detector performance, or radiation output, do not return the system to patient use until the failure is understood and resolved.

Provide another verified mobile X-ray unit when needed.

Expected outcome: A system with unresolved performance concerns is not used clinically.

2. Confirm Which Test Failed

Identify the exact calibration or QC procedure, the stage where it failed, the displayed message, and whether the test previously passed.

Do not substitute a different test for the failed requirement.

Expected outcome: The specific failure is documented and reproducible.

3. Verify Test Setup

Confirm the approved test object, detector, positioning, orientation, distance, and other setup elements are correct according to the applicable procedure.

Do not invent or approximate test geometry.

Expected outcome: The calibration or QC procedure is being performed with the correct setup. If setup error caused the failure, repeat the approved test and stop when it passes.

4. Inspect Detector and Imaging Surfaces

Check the detector and test object for contamination, damage, improper covers, or foreign material that could affect the measurement.

Expected outcome: Imaging surfaces and test objects are clean and intact.

5. Confirm System Readiness and Power

Verify the GM85 Fit has completed startup, the intended detector is ready, battery state is adequate, and no other system warning is active.

Expected outcome: The system is stable and ready before the test begins.

6. Check Environmental Conditions

Look for unusual temperature conditions, vibration, direct interference from nearby equipment, unstable flooring, or other obvious environmental changes that could affect testing.

Expected outcome: Testing is performed under appropriate and stable environmental conditions.

7. Verify Test Equipment

If external meters, phantoms, fixtures, or accessories are used, confirm they are the correct equipment and their calibration or verification status is current according to facility policy.

Expected outcome: The measurement equipment itself is not the source of the failure.

8. Repeat the Approved Test Once Conditions Are Correct

After correcting any external setup problem, repeat the applicable manufacturer or facility-approved test.

Do not repeatedly recalibrate the system in an attempt to force a passing result.

Expected outcome: The system passes the required test. If the failure repeats under correct conditions, stop external troubleshooting and escalate.

9. Document Return-to-Service Evidence

Record the test performed and passing result according to department policy before releasing the unit.

Expected outcome: Objective evidence supports return to clinical use.

If the Problem Persists

If setup, detector condition, system readiness, environment, and test equipment have been verified and the QC or calibration still fails, common external causes have been ruled out.

Possible remaining categories include detector calibration, generator performance, image-processing calibration, internal sensors, acquisition electronics, software, or service-level adjustment requirements.

The system should be:

Required QC, calibration, radiation-performance, and functional testing must pass before return to service.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Never treat a failed quality-control test as a nuisance alarm; determine whether it affects confidence in patient imaging before returning the system to use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiology reported that the GM85 Fit detector quality-control test would not complete successfully."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the detector was positioned incorrectly for the approved QC setup."

Resolution

What action was taken.

Example:
"Clinical Engineering corrected the detector and test-object positioning, repeated the approved QC procedure successfully, and documented the passing result."

Helpful Details to Include (If Known)

Final Thought

A failed QC test should be approached methodically by verifying setup and test conditions before assuming equipment failure, while preserving objective evidence and escalating any persistent performance concern.

That is successful troubleshooting.