What This Guide Helps With

Failure to send images to PACS, DICOM transfer errors, network disconnects, or studies not appearing in the archive.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient care continues without interruption
Why it matters: Export issues should never impact clinical workflow or diagnosis

Verify the Reported Issue

Expected: Issue is reproducible
Why it matters: Confirms this is a true communication failure, not user workflow error

Check Network Cable Connection

Expected: Solid or blinking network lights
Why it matters: Loose or disconnected cables are a common cause of DICOM failure

Verify Network Connectivity

Expected: Valid IP address and network connection
Why it matters: No network = no DICOM communication

Confirm DICOM Configuration Settings

Expected: All values match hospital PACS configuration
Why it matters: Incorrect configuration is one of the most common causes of failed exports

Perform a Test Send

Expected: Study appears in PACS
Why it matters: Confirms end-to-end communication is working

Check for Network Restrictions

Expected: No network-level blocks
Why it matters: Even correctly configured systems will fail if blocked by network security

Restart the System

Expected: Temporary communication issues may resolve
Why it matters: Clears software/network stack issues

Try Alternate Network Port

Expected: Successful connection and transfer
Why it matters: Rules out faulty wall ports or switch issues

If the Problem Persists

If all external connections, network settings, and configuration have been verified, the issue is likely internal (network interface hardware or system software).

Remove the device from service, label it Out of Service, and send for repair or further bench evaluation.

Knowing when to stop prevents unnecessary downtime and protects system integrity.

Clinical Use Tip

Always confirm images have successfully reached PACS before ending the exam. Do not troubleshoot on an active patient. Move the patient to a backup device first to ensure therapy continuity.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ultrasound unable to send studies to PACS; export failing."

Cause

What was observed during troubleshooting.

Example:
"Incorrect DICOM AE Title configured, preventing communication with PACS server."

Resolution

What action was taken.

Example:
"Corrected AE Title and verified successful test transmission to PACS."

Helpful Details to Include (If Known)

Final Thought

DICOM failures are most often caused by configuration or network issues, not hardware failure. A structured approach—starting with connections and settings—ensures safe, efficient resolution while minimizing downtime. Proper escalation and documentation support long-term system reliability.

That is successful troubleshooting.