Samsung Healthcare HM70A

Image Dropout or Quality Degradation

On this page

Asset Type

Ultrasound System

Manufacturer

Samsung Healthcare

Model

HM70A

What This Guide Helps With

Intermittent or poor ultrasound image quality, including dropout, noise, or weak signal caused by probe, connection, or system setting issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Issue

Check Probe Connection (Most Common Cause)

Test with a Known Good Probe

Inspect the Probe Cable

Check Imaging Settings

Check for Overheating or Extended Use Effects

Inspect Probe Port for Damage

Perform a System Reboot

If the Problem Persists

If all probes show poor image quality or dropout after these checks, common external causes have been ruled out.

The issue is likely internal (beamformer, processing board, or port hardware).

Remove the system from service, label as Out of Service, and send for vendor repair or bench evaluation.

Clinical Use Tip

Never rely on a degraded ultrasound image for diagnosis. Always transition to a known-good system before continuing patient care.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Ultrasound image intermittently dropping out during use on Samsung HM70A."

Cause

What was observed during troubleshooting.

Example:
"Faulty linear probe cable causing intermittent signal loss when manipulated."

Resolution

What action was taken.

Example:
"Replaced defective probe with known good unit; verified normal image quality restored."

Helpful Details to Include (If Known)

Final Thought

Image quality issues should always be approached from the outside in—probe, cable, and connection first. Most failures are not internal. Protect patient safety by escalating when needed and documenting clearly.

That is successful troubleshooting.