Schiller CARDIOVIT AT Series

Battery Not Charging

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

Schiller

Model

CARDIOVIT AT Series

What This Guide Helps With

EKG not charging, running only on battery, or shutting down due to power source, battery, or connection issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient monitoring continues uninterrupted.

Why it matters: Power loss can interrupt ECG acquisition and documentation.

Verify Wall Outlet Power

Expected: Outlet provides stable power.

Why it matters: A dead or switched outlet is a common external cause.

Inspect AC Power Cord and Adapter

Expected: Secure, undamaged connection.

Why it matters: Faulty cords or adapters prevent charging.

Confirm AC Power Indicator on Device

Expected: AC indicator is lit and device recognizes external power.

Why it matters: Confirms whether the device is receiving power.

Check Battery Charging Status on Display

Expected: Battery shows charging status.

Why it matters: Helps determine if the issue is battery vs charging circuit.

Perform a Power Reset

Expected: Charging resumes after reboot.

Why it matters: Temporary faults can prevent charging recognition.

Check for Docking Station or External Charger (If Applicable)

Expected: Proper contact allows charging.

Why it matters: Poor contact prevents current flow to the battery.

Inspect for Environmental or Physical Issues

Expected: No abnormal conditions present.

Why it matters: Environmental damage can affect charging circuits.

If the Problem Persists

If all external power sources, cables, and connections have been verified, the issue is likely internal (battery failure, charging circuit, or power board).

Remove the device from service, label it Out of Service, and send it for repair or bench evaluation.

Knowing when to stop prevents unnecessary risk and saves troubleshooting time.

Clinical Use Tip

Do not troubleshoot charging issues while the device is actively needed for patient care. Always move the patient to a functioning EKG machine first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"EKG machine not charging and shutting down during use."

Cause

What was observed during troubleshooting.

Example:
"Battery failed to accept charge after verifying power cord, outlet, and adapter were functional."

Resolution

What action was taken.

Example:
"Removed device from service and sent to Clinical Engineering for battery replacement and evaluation."

Helpful Details to Include (If Known)

Final Thought

Effective troubleshooting starts with simple external checks and confirms power delivery before assuming internal failure. Patient safety must always come first, and proper escalation ensures reliable equipment performance.

That is successful troubleshooting.