Schiller CARDIOVIT AT Series

Network or EMR Connectivity Failure

On this page

Asset Type

Electrocardiograph (EKG) Machine

Manufacturer

Schiller

Model

CARDIOVIT AT Series

What This Guide Helps With

Failure to transmit ECGs to EMR, network disconnects, or missing studies due to cable, Wi-Fi, or configuration issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient diagnostics continue without interruption.

Why it matters: Network issues should not delay clinical decision-making.

Verify the Reported Issue

Expected: Clear understanding of the failure type.

Why it matters: Helps determine if issue is network, configuration, or workflow-related.

Check Physical Network Connection (Wired Systems)

Expected: Link lights active and stable connection.

Why it matters: Loose or failed cables are a common cause.

Verify Wi-Fi Connection (If Applicable)

Expected: Device shows connected status with adequate signal.

Why it matters: Weak or dropped Wi-Fi prevents data transmission.

Confirm Network Configuration Settings

Expected: Valid network configuration present.

Why it matters: Incorrect settings prevent communication on the network.

Check EMR / Server Configuration Settings

Expected: Settings match hospital configuration.

Why it matters: Misconfiguration prevents successful export even with network connection.

Attempt Test Transmission

Expected: Successful transmission without errors.

Why it matters: Confirms whether issue is resolved or persistent.

Check for Network-Wide Issues

Expected: Identify if issue is isolated or system-wide.

Why it matters: Helps differentiate device issue from infrastructure problem.

Power Cycle the Device

Expected: Network reconnects and export resumes.

Why it matters: Clears temporary communication or software faults.

If the Problem Persists

All common external and configuration-related causes have been ruled out. The issue is likely internal (network interface, mainboard, or software fault).

Remove the device from service, label it Out of Service, and send for repair or further bench evaluation.

Knowing when to stop prevents unnecessary risk and wasted troubleshooting time.

Clinical Use Tip

Always complete ECG acquisition first. If the network is down, print or manually upload results to avoid delays in patient care.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"EKG not transmitting studies to EMR; staff reports export failures."

Cause

What was observed during troubleshooting.

Example:
"Misconfigured network settings with incorrect gateway address."

Resolution

What action was taken.

Example:
"Corrected network configuration and verified successful test transmission."

Helpful Details to Include (If Known)

Final Thought

Effective troubleshooting starts with simple external checks like cables and network status before assuming device failure. Patient care must always come first, and proper escalation ensures safety and efficiency. Clear documentation supports continuity and future troubleshooting.

That is successful troubleshooting.