On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots data-transfer failures involving network availability, wireless signal, configuration, destination systems, credentials, or infrastructure.
Step-by-Step Troubleshooting
1. Protect Clinical Care and Documentation
Do not delay emergency treatment because electronic data transfer is unavailable.
Ensure patient information, ECG records, and treatment documentation are preserved using an approved alternate workflow while troubleshooting occurs.
Expected outcome: Patient care and required clinical documentation continue without dependence on the failed data connection.
2. Confirm the Exact Transfer Failure
Determine whether the problem affects:
- Wi-Fi connection.
- Cellular connection.
- ePCR transfer.
- SentioWeb communication.
- One case only.
- All cases.
- One destination or all destinations.
Record any displayed network or transfer message.
Expected outcome: The affected communication path is clearly identified.
3. Confirm the Device Is Otherwise Operating Normally
Verify the DEFIGARD Touch 7 starts normally and that patient monitoring and therapy functions are not affected.
If the device exhibits broader instability, remove it from service rather than treating the problem as communications-only.
Expected outcome: The failure is isolated to connectivity or data transfer.
4. Check Wireless Signal and Environment
Move the device, when clinically appropriate, to an area where reliable facility Wi-Fi or cellular service is normally available.
Check for:
- Poor coverage.
- Shielded rooms.
- Basements.
- Elevators.
- Temporary network outages.
- Known cellular dead zones.
Expected outcome: The device has adequate wireless coverage. If transfer succeeds in a known-good coverage area, document an environmental or infrastructure-related issue.
5. Verify User-Accessible Network Status
Review normal status indications for Wi-Fi or cellular connectivity without altering protected network settings.
Confirm whether the device is:
- Connected.
- Searching.
- Disconnected.
- Connected locally but unable to reach the destination.
Expected outcome: Network status is clearly established.
6. Compare With Another Device or Location
Determine whether another configured DEFIGARD Touch 7 or compatible hospital device can communicate from the same location.
If multiple devices fail simultaneously, involve the appropriate network, cellular, ePCR, or application support team.
Expected outcome: The problem is isolated to the individual device or identified as an infrastructure/service issue.
7. Check Data Workflow and Destination Availability
Confirm that the intended ePCR or SentioWeb destination is available and that the reported record has been finalized or made ready for transfer according to local workflow.
Do not alter protected credentials or destination configuration without authorization.
Expected outcome: The data is in a transferable state and the intended receiving service is available.
8. Retry With a Test or Approved Nonclinical Record
When permitted by policy, perform a controlled data-transfer test using an approved test record or workflow.
Expected outcome: Data reaches the intended destination successfully. If transfer is restored and repeatable, troubleshooting can stop.
9. Perform Final Verification
Confirm the device reconnects normally after restart or movement between expected coverage areas, if applicable, and that no clinically required records remain unsent.
Expected outcome: Communication is stable and expected data is transferred or otherwise accounted for.
If the Problem Persists
Common coverage, workflow, destination, and external infrastructure causes have been ruled out. The remaining issue may involve device communications hardware, configuration, certificates, application integration, network authorization, or remote service infrastructure.
The device should be removed from service when communications are required for local clinical operations or when device reliability is uncertain. Otherwise, follow local policy for restricted use.
For unresolved failures:
- Label the affected function or device Out of Service as appropriate.
- Send the device for bench evaluation when indicated.
- Engage hospital IT, integration, ePCR, cellular, or manufacturer support as appropriate.
- Use approved manufacturer documentation and network tools.
- Allow configuration changes only by authorized personnel.
Verify data transfer before restoring the affected workflow.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Confirm the complete path from the defibrillator to the receiving ePCR or data system; a wireless connection icon alone does not prove successful record delivery.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Staff reported that treatment records from the DEFIGARD Touch 7 were not transferring to the ePCR system over Wi-Fi."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found that the unit was being used in an area with inadequate wireless coverage while identical devices transferred normally elsewhere."
Resolution
What action was taken.
Example:
"Verified successful transfer from a known-good coverage area and referred the location-specific wireless issue to the appropriate network support team."
Helpful Details to Include (If Known)
- Connection method involved.
- Exact error or status message.
- Device location.
- Signal or connection indication.
- Whether other devices are affected.
- Whether one or all records fail.
- Destination system involved.
- Test-transfer result.
- IT or application ticket number.
- Final communication status.
Final Thought
Keep treatment and documentation moving while isolating whether the failure belongs to the device, wireless environment, workflow, or receiving infrastructure. Avoid unauthorized network changes and document the complete communication-path test before escalation.
That is successful troubleshooting.