What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting situations where the ACUSON Sequoia ultrasound system fails to recognize or detect a connected probe. Common presentations include:

This guide focuses on simple, external checks before considering internal repair.

Step-by-Step Troubleshooting

Verify Probe Connection

Inspect the Probe Cable

Swap Ports (If Applicable)

Test with a Known-Good Probe

Check System Software and Firmware

Restart the System

Document Observations

If the Problem Persists

External issues have likely been ruled out.

Knowing when to stop is proper troubleshooting—do not attempt internal electronics repair.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“No Probe Connected” message on the ACUSON Sequoia system.

Cause

What was observed during troubleshooting.

Example:
Probe cable or connector may be loose or damaged; system may not be recognizing the probe.

Resolution

What action was taken.

Example:
Verified connections and cable integrity, swapped probe port, tested with known-good probe, restarted system. If problem persisted, system and/or probe labeled Out of Service and sent for service evaluation.

Helpful Details to Include

Final Thought

Systematic, stepwise troubleshooting ensures patient safety while preserving equipment integrity. By checking external connections first, verifying software, and testing known-good probes, Clinical Engineering can identify whether the issue is user-serviceable or requires escalation. Proper documentation using CCR reinforces accountability and efficiency.

That is successful troubleshooting.