Siemens Healthineers Cios Alpha

DICOM/PACS Export Failure

On this page

Asset Type

C-Arm

Manufacturer

Siemens Healthineers

Model

Cios Alpha

What This Guide Helps With

Troubleshooting Cios Alpha DICOM or PACS export failures caused by network, destination, patient data, or configuration issues.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

Before troubleshooting, confirm the C-arm is not actively needed for a procedure.

Action:

Expected outcome: Patient care continues safely, and stored images are protected.

Why it matters: A DICOM/PACS export issue is usually a connectivity or workflow issue, but accidental deletion or interruption during a case can create a patient care risk.

2. Verify the Reported Export Failure

Confirm exactly what is failing.

Action:

Expected outcome: You identify whether this is a single-study issue, destination issue, or system-wide export problem.

If export works for other studies or destinations, stop broad troubleshooting and focus on the affected study, patient data, or destination.

3. Confirm the C-Arm Is Connected to the Network

Check the simplest external network causes first.

Action:

Expected outcome: The C-arm has a valid physical network connection.

If a loose or disconnected cable is found, reconnect it and retry export. If the export succeeds, document the cable/network connection issue and stop.

4. Test With a Known-Good Network Drop or Cable

Rule out a bad cable, jack, or local network path.

Action:

Expected outcome: The export either works on a known-good connection or continues to fail.

If the export works after changing the cable or network drop, the original cable, jack, or network path is the likely cause. Stop and document the finding.

5. Check for Local Network Address Problems

Confirm the system appears to have normal network connectivity.

Action:

Expected outcome: The C-arm has a valid network identity for the imaging network.

If the IP address is missing, incorrect, duplicated, or outside the expected range, escalate to IT/PACS support or approved service support before changing network configuration.

6. Confirm the Correct DICOM/PACS Destination Is Selected

A wrong destination can look like a device failure.

Action:

Expected outcome: The system is attempting export to the correct destination.

If another DICOM destination works, the issue is likely destination-specific rather than a total C-arm failure.

7. Verify Patient and Study Information

Incorrect or incomplete patient/study data can interfere with DICOM workflow.

Action:

Expected outcome: Required DICOM fields are complete and reasonable.

If correcting patient/study information allows export, document the data issue and stop.

8. Check Modality Worklist or Order Association

If the site uses modality worklist, export may depend on proper order selection.

Action:

Expected outcome: The exported study is properly associated with the correct order.

If the study exports after proper order association, the issue was likely workflow or demographic mismatch.

9. Review the Export Queue

Check whether studies are stuck, pending, failed, or repeatedly retrying.

Action:

Expected outcome: You determine whether the problem is a queue backlog, a single failed study, or total DICOM communication failure.

If one study is blocking the queue, escalate to PACS/IT or vendor support before deleting or altering data.

10. Confirm PACS or DICOM Destination Availability

The C-arm may be functioning normally while the receiving system is unavailable.

Action:

Expected outcome: The receiving PACS/DICOM node is confirmed available.

If other devices also cannot export, the issue is likely PACS, network, or destination-side rather than the Cios Alpha.

11. Check for Recent Configuration Changes

Determine whether the issue started after a change.

Action:

Expected outcome: You identify whether the failure began after a known environmental or configuration change.

If the issue began after a network or PACS change, coordinate with IT/PACS and vendor support as needed.

12. Power Cycle Only When Safe

A restart may clear a temporary software or communication hang, but it should not interrupt care.

Action:

Expected outcome: Temporary software or network session issues may clear.

If export succeeds after restart, monitor for recurrence and document the restart as the corrective action.

13. Escalate for DICOM Configuration or Internal Fault Evaluation

Do not guess or make unauthorized DICOM configuration changes.

Action:

Expected outcome: The correct support team receives enough information to isolate the failure.

If the Problem Persists

If the Cios Alpha still cannot export after checking network connection, cable, network drop, destination selection, patient data, worklist association, export queue, and PACS availability, the issue is likely configuration-related, software-related, network-side, or internal to the system.

The device should be:

Knowing when to stop is proper troubleshooting. DICOM and PACS problems often involve both the medical device and hospital network environment, so escalation should include accurate details rather than unnecessary internal disassembly.

Clinical Use Tip

Do not troubleshoot export workflow during an active procedure unless the clinical team confirms it is safe. If imaging is needed, allow the case to continue locally and resolve the PACS export issue after patient care is protected. Do not troubleshoot on an active patient when therapy or imaging continuity is at risk; move patient care to a backup device or workflow first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Siemens Healthineers Cios Alpha C-arm would not export images to PACS after a completed case."

Cause

What was observed during troubleshooting.

Example:
"Troubleshooting found the C-arm connected to a network wall jack that was not configured for the imaging/PACS network."

Resolution

What action was taken.

Example:
"Reconnected the C-arm to an approved imaging network port, verified local image storage, successfully exported a test study to PACS, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

DICOM/PACS export failures should be approached logically. Confirm patient safety first, then verify local image availability, network connection, destination selection, patient demographics, worklist association, and PACS availability. Avoid unnecessary internal repair assumptions until external network and workflow causes are ruled out. Clear CCR documentation helps Clinical Engineering, PACS, IT, and vendor support resolve the issue faster.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide