Siemens Healthineers ACUSON Juniper

DICOM send, worklist, network, or export failure

On this page

Asset Type

Ultrasound System

Manufacturer

Siemens Healthineers

Model

ACUSON Juniper

What This Guide Helps With

Troubleshooting ultrasound DICOM send, modality worklist, network connection, archive, USB, or image export failures before repair escalation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected outcome: Patient images and exam data are protected before troubleshooting begins.

Confirm the Exact Communication or Export Failure

Expected outcome: The issue is separated into DICOM, worklist, general network, or removable media export failure.

Check Whether the Exam Is Still Stored Locally

Expected outcome: Clinical data is preserved while the transfer problem is investigated.

Check Physical Network Connection

Expected outcome: Loose cable, damaged cable, or inactive wall port is ruled out.

Check Network Link Indicators

Expected outcome: Basic network link status is confirmed before changing system configuration.

Confirm the System Is on the Correct Network

Expected outcome: The failure is not caused by using a non-imaging network port or unauthorized network connection.

Check IP Address and Network Configuration

Expected outcome: Incorrect IP configuration or failed address assignment is identified.

Check for Duplicate or Held IP Address Concerns

Expected outcome: Network-side assignment issues are ruled out before suspecting the ultrasound system.

Test Basic Network Connectivity

Expected outcome: The issue is narrowed to local network access, DICOM destination communication, or server-side availability.

Verify DICOM Destination Settings

Expected outcome: Incorrect DICOM destination settings are corrected or ruled out.

Confirm PACS or Archive Availability

Expected outcome: System-specific failure is separated from wider PACS/network downtime.

Verify Modality Worklist Settings

Expected outcome: Worklist failure caused by incorrect query or server configuration is identified.

Check Patient Query Workflow

Expected outcome: User workflow or scheduling mismatch is ruled out before hardware escalation.

Try a Test Send or Known Good Study

Expected outcome: Transfer behavior helps identify whether the failure is local, network-related, or PACS-side.

Check the DICOM Queue or Transfer Status

Expected outcome: The failure is documented with enough detail for IT, PACS, or vendor escalation.

Check System Date and Time

Expected outcome: Date and time mismatch is ruled out.

Restart Network-Related Workflow Safely

Expected outcome: Temporary software or communication lockup is cleared.

Check USB Export Media

Expected outcome: Failed removable media is ruled out.

Check USB Port Condition

Expected outcome: A port-specific issue is identified without internal disassembly.

Confirm Export Format and Destination

Expected outcome: Export failure due to format, destination, or storage space is ruled out.

Check Local Storage Status

Expected outcome: Storage-related communication or export problems are identified safely.

Review Recent Changes

Expected outcome: The issue is linked to a recent change when applicable.

Coordinate With IT or PACS Support

Expected outcome: External network and server-side causes are properly investigated.

Escalate If Communication Hardware or Software Fault Is Suspected

Expected outcome: The device is escalated for bench evaluation or vendor service rather than repeated ineffective troubleshooting.

If the Problem Persists

If the problem continues after checking cabling, network configuration, DICOM settings, worklist settings, PACS availability, export media, USB port condition, and local storage, common external causes have been ruled out.

The issue may involve the ultrasound system network adapter, software services, internal storage, DICOM application configuration, USB hardware path, or deeper system fault.

The device should be:

Knowing when to stop is proper troubleshooting. Repeated configuration changes without confirming the network and PACS side can create additional problems.

Clinical Use Tip

Do not troubleshoot DICOM, worklist, network, or export failures during an active patient exam unless the device is safe, the patient is not at risk, and the exam data is protected. If image transfer is unreliable, make sure the department has a safe process to preserve studies locally or move the patient to another ultrasound system.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sonography reported the Siemens ACUSON Juniper could not send completed studies to PACS and the DICOM queue showed repeated failed transfers."

Cause

What was observed during troubleshooting.

Example:
"Verified local studies were saved, Ethernet cable was connected, but the ultrasound was connected to a network jack not configured for imaging DICOM traffic."

Resolution

What action was taken.

Example:
"Moved the system to an approved imaging network port, confirmed worklist query loaded, test study sent successfully to PACS, and returned the unit to service."

Helpful Details to Include (If Known)

Final Thought

DICOM, worklist, network, and export failures should be approached logically from the outside in. Protect patient data first, confirm basic network and media conditions, verify configuration with IT or PACS, and escalate when the issue points beyond safe Clinical Engineering checks. Clear CCR documentation helps prevent repeated troubleshooting and supports faster resolution.

That is successful troubleshooting.