On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Use this guide when a reagent is not recognized, shows incorrect status, or does not match the analyzer's expected onboard inventory.
Step-by-Step Troubleshooting
1. Protect Patient Testing and Confirm the Reagent Issue
Do not release patient results from an assay when the required reagent identity, status, or availability cannot be verified.
Determine whether the issue involves:
- One reagent
- One assay
- Multiple reagent positions
- Incorrect reagent quantity or status
- Reagent physically present but not recognized
- Inventory changing unexpectedly
Document the exact reagent and displayed condition.
Expected outcome: The affected reagent and scope of the inventory problem are clearly identified. If the issue is corrected and reagent status becomes reliable, troubleshooting can stop after required verification.
2. Verify the Correct Reagent Is Present
Confirm laboratory staff loaded the intended reagent for the assay and analyzer.
Inspect for:
- Wrong reagent container
- Wrong assay material
- Duplicate or swapped containers
- Incorrect loading location
- Reagent not fully seated
- Packaging or protective material left in place
Do not alter reagent contents or bypass normal reagent-identification controls.
Expected outcome: The correct reagent is installed in the intended location. If proper loading restores recognition, the issue is resolved.
3. Inspect the Reagent Container and Identification Features
Inspect the reagent container externally for:
- Damage
- Leakage
- Contamination
- Damaged identification label
- Wrinkled or obscured machine-readable information
- Debris on surfaces involved in recognition
- Improper closure or assembly
Remove leaking or damaged reagents from use according to laboratory procedure.
Expected outcome: The reagent container is intact and its identification features are readable. If replacing a damaged container restores recognition, troubleshooting can stop after verification.
4. Remove and Reload the Reagent Correctly
If permitted by laboratory procedure, remove and reload the affected reagent using the normal loading process.
Confirm:
- Correct orientation
- Full seating
- No obstruction
- Loading area is clean
- Analyzer acknowledges the load event
Avoid repeated removal and reinsertion that could compromise reagent handling.
Expected outcome: The analyzer recognizes the reagent and displays the expected identity and usable status. If so, the issue is resolved.
5. Compare With a Known-Good Reagent
Where permitted, test with another valid reagent of the same intended type or compare behavior with a known-good reagent loaded in another suitable position.
Determine whether the issue follows:
- The reagent container
- A specific reagent position
- A broader analyzer inventory problem
Expected outcome: The source is narrowed to the reagent or analyzer position. If the problem follows one reagent container, remove that reagent from use and stop troubleshooting once normal recognition is confirmed.
6. Verify Inventory Display Against Physical Inventory
Compare the analyzer's displayed reagent inventory with the physical reagents actually present.
Check for:
- Reagent shown onboard when it has been removed
- Reagent physically present but absent from inventory
- Incorrect position assignment
- Unexpected quantity or availability status
- Disagreement affecting one assay versus many assays
Do not manually alter inventory through unauthorized service functions.
Expected outcome: Displayed inventory agrees with the physical analyzer state. If normal operation is restored through approved unload/reload procedures, the issue is resolved.
7. Inspect Accessible Reagent Loading Areas
With the analyzer in an appropriate safe state, inspect accessible reagent loading areas for:
- Spills
- Residue
- Debris
- Foreign objects
- Mispositioned containers
Clean only surfaces permitted under normal maintenance procedures.
Expected outcome: No accessible contamination or obstruction interferes with reagent recognition. If approved cleaning corrects the issue, verify recognition and stop troubleshooting.
8. Verify the Analyzer Is Otherwise Ready
Check whether a larger analyzer condition is affecting reagent inventory, such as:
- Incomplete initialization
- Active communication fault
- Reagent subsystem unavailable
- Software or interface not updating
- Other reagent positions also reporting incorrectly
Expected outcome: The analyzer is otherwise operational and the reagent problem is either resolved or clearly isolated.
9. Perform Final Verification
After correction, confirm:
- Correct reagent identity
- Correct position
- Appropriate analyzer status
- Assay availability
- Required calibration and quality-control status before patient testing
Expected outcome: Reagent inventory is stable and the associated assay is ready according to laboratory requirements. Troubleshooting can stop.
If the Problem Persists
If the reagent is correct, intact, properly loaded, and recognized inconsistently across known-good substitutions, common external causes have been ruled out.
Possible remaining categories include an internal reagent-identification reader, position sensor, inventory database or software problem, communications issue, or other service-level condition.
The analyzer should be:
- Removed from service for affected testing when reagent identity cannot be assured
- Labeled Out of Service when the problem compromises analyzer reliability
- Sent for qualified repair or service evaluation
- Evaluated using Siemens Healthineers documentation and approved test equipment
- Repaired or configured only by qualified personnel
Complete appropriate reagent, calibration, QC, and functional verification before returning affected assays to patient testing.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Do not override or work around uncertain reagent identification; reagent identity and status must be reliable before reporting patient results.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Laboratory staff reported that one Atellica IM 1600 reagent was physically loaded but did not appear correctly in onboard inventory."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the reagent container was not fully seated and confirmed the loading position and analyzer were otherwise normal."
Resolution
What action was taken.
Example:
"The reagent was reloaded correctly, inventory updated to the expected status, and assay readiness was verified before returning the analyzer to use."
Helpful Details to Include (If Known)
- Assay and reagent affected
- Reagent identity and physical condition
- Displayed inventory status
- Loading position
- Whether unload/reload corrected the issue
- Known-good reagent comparison
- Evidence of spills or debris
- Other reagent positions affected
- Calibration or QC status after correction
- Final analyzer status
Final Thought
Reagent problems should be approached by confirming the correct material, physical condition, seating, identification, and inventory agreement before considering an analyzer fault. Patient testing should resume only when reagent status is dependable and verified.
That is successful troubleshooting.