Siemens Healthineers Atellica IM 1600

Result Is Delayed, Missing, or Inconsistent

On this page

Asset Type

Immunoassay Analyzer

Manufacturer

Siemens Healthineers

Model

Atellica IM 1600

What This Guide Helps With

Use this guide when expected results are delayed, absent, unexpectedly inconsistent, or do not follow the normal sample-to-result workflow.

Step-by-Step Troubleshooting

1. Protect Patient Care and Confirm the Exact Result Problem

Do not release a questionable result or wait indefinitely on an analyzer when patient care depends on timely testing.

Identify whether the result is:

If result integrity is uncertain, involve laboratory personnel and use an alternate validated testing path as appropriate.

Expected outcome: The issue is categorized as processing delay, missing transmission, absent result, or questionable analytical result.

2. Confirm the Correct Sample and Order

Verify:

Expected outcome: The analyzer is working on the intended specimen and test. If the problem was an order or identification mismatch, correct it through the approved laboratory process and verify the result pathway.

3. Check Analyzer and Sample Status

Determine where the specimen currently resides in the workflow.

Look for:

Expected outcome: The sample's processing state is known and any external condition delaying it is identified.

4. Verify Reagent, Calibration, and QC Readiness

For the affected assay, confirm:

Do not bypass an assay hold caused by unresolved calibration or QC.

Expected outcome: The assay is properly qualified for testing. If correcting an identified reagent, calibration, or QC condition restores normal processing, troubleshooting can stop after verification.

5. Check for Sample-Handling Problems

Review whether the affected specimen experienced:

Inspect the sample and rack as appropriate.

Expected outcome: Sample handling is either confirmed normal or a specific cause of the delay is identified and corrected.

6. Distinguish Analyzer Result Generation From LIS Transmission

If the result appears correctly on the analyzer but not in the LIS, treat the problem as a communication or interface issue rather than an analytical failure.

Check for:

Expected outcome: Missing-result issues are correctly separated into analyzer-generation versus result-transmission problems.

7. Investigate Inconsistent Results Safely

For an unexpectedly inconsistent result, do not assume the analyzer is faulty based on one comparison.

Work with laboratory personnel to review:

Clinical Engineering should not make clinical interpretation decisions.

Expected outcome: The inconsistency is traced to a sample, assay, QC, or analyzer-level pattern rather than speculation.

8. Check for Broader Analyzer Delays

Determine whether:

Expected outcome: The problem is categorized as isolated or system-wide.

9. Verify the Complete Workflow After Correction

Using appropriate laboratory-approved material, verify:

Expected outcome: The complete sample-to-result pathway is reliable. Troubleshooting can stop.

10. Escalate Persistent Missing or Inconsistent Results

If sample identification, handling, reagent, calibration, QC, processing status, and communication have all been verified but results remain unreliable, remove the affected assay or analyzer from service.

Do not attempt internal detector adjustment, assay parameter changes, or restricted service calibration.

Expected outcome: Continued use of an unreliable analytical pathway is prevented and qualified support is engaged.

If the Problem Persists

If sample identity, sample condition, reagent availability, calibration, QC, analyzer status, and LIS communication have been verified, common external causes have been ruled out.

Possible remaining categories include internal measurement performance, fluidics, software processing, database or interface behavior, assay configuration, detector performance, or another service-level problem.

The affected analyzer or assay should be:

Before return to service, verify the full sample-to-result workflow and complete laboratory-required QC.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

For a missing result, trace the specimen from identification through processing and transmission; never assume “no result in the LIS” means the analyzer failed analytically.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory staff reported that an Atellica IM 1600 result was complete on the analyzer but remained missing from the LIS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the analyzer was processing normally but its network communication had been interrupted."

Resolution

What action was taken.

Example:
"Network communication was restored and an approved test confirmed successful result transmission from the analyzer to the LIS."

Helpful Details to Include (If Known)

Final Thought

Delayed, missing, and inconsistent results require tracing the complete analytical workflow rather than assuming an internal instrument failure. Protect patient-result integrity, verify every external dependency, and escalate when reliable performance cannot be demonstrated.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide