Siemens Healthineers Atellica Solution

Calibration, QC, Atellica Data Manager, or LIS Communication Failure

On this page

Asset Type

Clinical Chemistry Analyzer

Manufacturer

Siemens Healthineers

Model

Atellica Solution

What This Guide Helps With

Troubleshooting calibration, QC, Data Manager, host communication, or LIS result transmission issues using external and workflow checks first.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Atellica Solution is not actively processing urgent patient specimens affected by the reported issue.

If testing is delayed, misrouting, or producing questionable results, notify the lab to follow downtime or alternate analyzer workflow.

Expected outcome: Patient results are not released until calibration, QC, and communication status are verified.

Confirm the Exact Reported Failure

Ask laboratory staff what failed and when it started.

Identify whether the issue involves:

Expected outcome: The issue is separated into analytical, middleware, or LIS communication failure before troubleshooting.

Check Analyzer Status and Active Alarms

Review the analyzer screen for system status, module alarms, reagent flags, calibration alerts, QC alerts, sample handling errors, or communication warnings.

Do not clear alarms without recording them if they may explain the failure.

Expected outcome: The analyzer provides a clear starting point for whether the issue is assay-related, module-related, or communication-related.

Verify the Correct Assay, Lot, and Calibration Requirements

Confirm the affected assay uses the correct reagent lot, calibrator lot, QC lot, and active calibration status.

Check whether a new reagent lot, calibrator lot, QC lot, or assay pack was recently loaded.

Expected outcome: Lot mismatch, expired material, missing calibration, or incorrect assay setup is ruled out before suspecting the analyzer.

Check Reagent, Calibrator, and QC Material Condition

Confirm reagents, calibrators, and QC material are not expired, improperly stored, contaminated, insufficient, or loaded in the wrong position.

Have lab staff verify preparation, reconstitution, mixing, thawing, and storage requirements according to lab procedure.

Expected outcome: Material handling problems are ruled out as the cause of calibration or QC failure.

Repeat Only Appropriate Calibration or QC Workflow

If lab policy allows, have authorized lab staff repeat the failed calibration or QC after verifying materials and setup.

Clinical Engineering should not override failed QC or release patient results.

Expected outcome: A one-time handling or setup issue may be corrected without unnecessary service escalation.

Check for Related Sample, Probe, Fluidics, or Reagent Handling Errors

Review whether the calibration or QC failure is accompanied by aspiration errors, clot detection, low liquid detection, reagent recognition faults, wash errors, or fluidics alarms.

These issues can cause poor recovery or invalid calibration even when the assay setup is correct.

Expected outcome: The calibration or QC problem is linked to a possible analyzer handling issue if related alarms are present.

Verify Atellica Data Manager Availability

Confirm the Atellica Data Manager workstation or server is powered on, logged in as required, and not frozen or displaying application errors.

Check whether other connected analyzers or modules are communicating normally.

Expected outcome: A local Data Manager application or workstation issue is separated from the analyzer itself.

Check Network Connections

Inspect Ethernet cables at the analyzer, Data Manager workstation, network switch, and wall jack if accessible.

Confirm link lights are present where visible.

Do not move network cables to undocumented ports unless directed by IT or the lab interface team.

Expected outcome: Loose, disconnected, or visibly damaged network connections are ruled out.

Confirm Whether the Failure Is Analyzer-to-Data Manager or Data Manager-to-LIS

Determine where the communication stops.

Check whether:

Expected outcome: The communication failure is narrowed to the analyzer, middleware, network, or LIS side.

Check for Interface Queue, Held Results, or Rejected Results

Review Data Manager status for held, pending, rejected, or untransmitted results.

Confirm whether results are being held due to QC rules, delta checks, missing demographics, invalid accession numbers, or LIS rejection.

Expected outcome: Results are not mistaken as “not transmitting” when they are actually being held by rules or rejected by LIS.

Check Recent Changes

Ask whether there were recent changes to:

Expected outcome: A recent change is identified as a likely cause before hardware repair is pursued.

Power Cycle Only When Safe and Approved

If the analyzer or Data Manager appears frozen or communication is down, coordinate with the lab before rebooting anything.

Confirm no active testing, critical samples, or pending transmissions will be interrupted.

Expected outcome: Controlled restart is performed only when safe and appropriate.

Test with a Known Workflow

After basic checks, have lab staff test a known valid workflow, such as a QC run, test order download, or result transmission according to site procedure.

Expected outcome: The team confirms whether the issue is resolved or still present after external checks.

If the Issue Resolves, Document and Return to Service

If calibration passes, QC is acceptable, Data Manager communication resumes, or LIS transmission is restored, confirm with lab staff before returning the analyzer to normal use.

Expected outcome: The device is operational and lab staff agree the issue is resolved.

If the Problem Persists

If calibration, QC, Data Manager, or LIS communication failure persists after material checks, lot verification, alarm review, network inspection, and workflow confirmation, common external causes have been ruled out.

The analyzer or related middleware/interface should be:

Knowing when to stop is proper troubleshooting. Clinical Engineering should not bypass failed QC, force result release, or make undocumented interface changes.

Clinical Use Tip

Do not troubleshoot calibration, QC, or LIS failures while patient results are actively being released. If result accuracy or transmission is uncertain, have the lab use an approved alternate analyzer or downtime workflow until the issue is resolved. Move patient specimen testing to a backup device or approved downtime workflow first so therapy and result continuity are protected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Laboratory reported the Atellica Solution was failing QC on chemistry assays and results were not transmitting through Atellica Data Manager to the LIS."

Cause

What was observed during troubleshooting.

Example:
"Found active QC holds in Data Manager and verified analyzer network link was present with no loose external connections; no hardware fault confirmed during initial troubleshooting."

Resolution

What action was taken.

Example:
"Verified reagent and QC lot information with lab staff, confirmed Data Manager queue status, coordinated LIS/interface review, and left analyzer restricted from affected testing pending successful QC and transmission verification."

Helpful Details to Include (If Known)

Final Thought

Calibration, QC, middleware, and LIS issues should be approached carefully because they can affect patient result accuracy and reporting. Start with materials, lot setup, alarms, workflow, and network checks before escalating. Good documentation protects the patient, the lab, and the troubleshooting process.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide