Siemens Healthineers Atellica Solution

Probe Aspiration, Clot Detection, or Sample Volume Error

On this page

Asset Type

Clinical Chemistry Analyzer

Manufacturer

Siemens Healthineers

Model

Atellica Solution

What This Guide Helps With

Troubleshooting sample aspiration, clot detection, insufficient sample, probe blockage, or volume-related errors using safe external checks first.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm whether any patient samples were affected by the error.

Do not release questionable results from an aspiration, clot, or insufficient-volume event until the analyzer status and sample integrity are verified.

Expected outcome: Patient results are protected from sampling-related errors.

Confirm the Exact Error or Event

Review the analyzer message, event log, or operator screen.

Identify whether the issue involves:

Expected outcome: The issue is separated into a sample-specific problem, probe/fluidics problem, or system-wide sampling issue.

Check Whether the Problem Follows One Sample

Ask the lab user whether the error occurred on one patient sample, one rack, or many samples.

If only one tube is affected, inspect that tube before suspecting the analyzer.

Expected outcome: A single problematic specimen is not mistaken for analyzer failure.

Inspect the Sample Tube Externally

Verify the tube is the correct type, properly seated, and not tilted in the carrier or rack.

Check for:

Expected outcome: The analyzer can access an acceptable sample without obstruction or misidentification.

If correcting the tube, sample volume, or rack seating resolves the issue, document the finding and stop.

Verify Minimum Sample Volume and Dead Volume

Confirm that the sample has enough usable volume for the ordered tests, including analyzer dead volume requirements.

A tube may visually appear adequate but still fail if the probe cannot safely aspirate the required amount.

Expected outcome: Short sample conditions are ruled out before analyzer service is requested.

Try a Properly Prepared Known-Good Sample or Control

With lab approval, test using an acceptable known-good sample, control, or approved troubleshooting material.

If the known-good sample runs correctly, the issue is likely related to the original specimen.

Expected outcome: Sample quality is separated from analyzer sampling performance.

Check Rack, Carrier, and Tube Positioning

Inspect the sample rack, carrier, or automation track position for misalignment, damage, debris, or incorrect loading.

Confirm the sample tube is fully seated and positioned correctly for probe access.

Expected outcome: Mechanical access problems from loading or rack issues are ruled out.

Inspect the Probe Area Externally

Visually inspect the sample probe area without disassembly.

Look for:

Expected outcome: Obvious external probe or sampling-area problems are identified.

If the probe appears bent, damaged, or contacting objects abnormally, remove the analyzer or affected module from service and escalate.

Check for Probe Wash or Carryover-Related Issues

Confirm the probe wash area is not visibly blocked and that waste or wash fluid is not overflowing.

Sampling errors may occur if the probe is not being cleaned correctly between aspirations.

Expected outcome: External wash station problems are considered before assuming internal probe failure.

Verify Reagent, Water, and Waste Status

Check system status for required fluids, reagent availability, wash solution, water supply, and waste levels.

Low system fluids, full waste, or fluidics-related alerts can contribute to aspiration or probe errors.

Expected outcome: System supply conditions are normal.

Review Recent Maintenance or Operator Actions

Ask whether the issue started after:

Expected outcome: The timing of the failure helps identify a recent external cause.

Perform Approved Operator-Level Cleaning or Maintenance

If available to Clinical Engineering or lab staff under site policy, run the approved probe cleaning, sample probe wash, or maintenance routine from the user interface.

Do not manually force, bend, remove, or internally disassemble the probe assembly unless following approved service procedure.

Expected outcome: Minor residue or external contamination is cleared safely.

If the error clears after approved cleaning and repeat testing passes, return the analyzer to service per site policy and document the action.

Check Whether the Error Is Isolated to One Module or Analyzer Function

Determine whether errors occur only on chemistry sampling, only on certain assays, or across multiple sample paths.

If multiple unrelated samples fail repeatedly, suspect analyzer sampling, probe, fluidics, or liquid level detection issues.

Expected outcome: The failure scope is defined before escalation.

Power Cycle Only When Appropriate

If the analyzer is not processing active samples and lab leadership approves, perform a controlled restart using the site-approved shutdown and startup process.

Do not interrupt active testing or sample processing without lab approval.

Expected outcome: Temporary software or motion state issues are cleared without compromising active patient testing.

Repeat a Controlled Test After Corrections

After correcting sample, rack, fluid, or cleaning issues, repeat testing using an approved sample or control.

Confirm:

Expected outcome: The analyzer demonstrates reliable sample aspiration before being returned to normal use.

If the Problem Persists

If the error continues after sample quality, tube positioning, rack loading, fluids, waste, and approved cleaning have been checked, common external causes have been ruled out.

The issue may involve the sample probe, liquid level detection system, clot detection function, syringe/fluidics path, pressure monitoring, wash station, sample handler alignment, or internal control hardware.

The affected analyzer, module, or sample path should be:

Knowing when to stop is proper troubleshooting. Do not continue running patient samples through a system that repeatedly reports aspiration, clot detection, or sample volume errors.

Clinical Use Tip

Do not troubleshoot sampling errors while the analyzer is actively processing critical patient specimens unless the lab has confirmed it is safe to pause or redirect testing. Protect result integrity first, then troubleshoot the analyzer. If testing must continue, move testing to a backup analyzer or approved downtime workflow first so therapy continuity and result integrity are protected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Lab reported the Siemens Atellica Solution was generating repeated sample aspiration and clot detection errors during chemistry testing."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found one affected specimen had low usable volume with visible fibrin, while rack seating and analyzer fluids appeared normal."

Resolution

What action was taken.

Example:
"Sample was removed for lab review, a properly prepared control/sample ran without aspiration errors, and the analyzer was returned to service with no repeat fault observed."

Helpful Details to Include (If Known)

Final Thought

Probe aspiration and sample volume errors should be approached logically from the outside in. Start with patient safety, specimen quality, tube position, rack loading, and fluid status before suspecting internal analyzer failure. If the issue repeats after external causes are ruled out, remove the affected system from service and escalate with clear CCR documentation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide