Siemens Healthineers Biograph Horizon

Patient Table Indexing, Movement, or Scan Range Positioning Issue

On this page

Asset Type

PET / CT System

Manufacturer

Siemens Healthineers

Model

Biograph Horizon

What This Guide Helps With

Troubleshooting patient table movement, indexing, scan range, table position, or bed travel issues using safe external checks before repair escalation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm whether a patient is on the PET / CT table.

If the table is not moving correctly, indexing unexpectedly, stopping during travel, or not aligning to the planned scan range, stop table movement immediately.

Expected outcome: The patient is protected from unexpected motion, incomplete positioning, or unsafe table travel.

If patient safety is affected, remove the patient from the system using the site’s approved procedure before continuing.

Confirm the Exact Table or Positioning Issue

Ask the technologist what occurred and when it occurred.

Identify whether the issue involves:

Expected outcome: The reported symptom is clearly defined before troubleshooting continues.

Check for Patient, Accessory, or Room Obstruction

Inspect the table travel path for anything blocking movement.

Check for:

Expected outcome: The table has a clear travel path with no external obstruction.

If removing an obstruction restores normal table movement, verify operation with staff and stop.

Verify the Patient Is Properly Positioned

Confirm the patient is centered and positioned according to the planned exam.

Check that arms, legs, blankets, immobilization aids, and positioning accessories do not extend into areas that may interfere with gantry clearance or table travel.

Expected outcome: Patient positioning does not create a physical movement limitation or scan range error.

Confirm Table Load and Weight Distribution

Verify that the patient and accessories do not exceed the site-approved table loading limits.

Also check that the patient is not positioned unusually far off-center or unsupported.

Expected outcome: The table is not being asked to move with excessive or poorly distributed load.

If table movement problems only occur with heavier patients or unusual positioning, remove the system from service until evaluated.

Check the Table Control Inputs

Test table movement using the normal operator controls when safe.

Check whether the issue occurs from:

Expected outcome: Clinical Engineering can determine whether the problem follows one control location or affects all table movement.

If only one external control is affected, inspect that control for damage, stuck buttons, contamination, or cable strain.

Inspect Emergency Stop and Safety Stop Conditions

Verify that emergency stop buttons are not engaged.

Check for safety stop, collision, or table interlock messages at the console or gantry display.

Expected outcome: Table motion is not being intentionally inhibited by a safety stop or interlock.

If an emergency stop was activated, do not simply reset and return to use until the reason for activation is understood.

Review System Messages at the Console

Ask staff to show the exact message related to table motion, indexing, scan range, positioning, or acquisition setup.

Record whether the message appears during:

Expected outcome: The fault timing helps separate workflow setup issues from actual table hardware or position feedback issues.

Confirm Scan Range Planning

Review the planned scan range with the technologist.

Confirm that the topogram/scout image, scan start location, scan end location, PET bed positions, and selected protocol match the intended anatomy.

Expected outcome: The reported scan range issue is not caused by incorrect planning, wrong protocol selection, or incorrect landmarking.

If correcting the planned range resolves the issue, document as a workflow/setup finding and stop.

Verify Patient Landmarking and Table Reference Position

Confirm that the patient was landmarked correctly before the scan.

Check whether the table was moved after landmarking, whether the wrong anatomical reference was selected, or whether the exam was planned from an outdated scout/topogram.

Expected outcome: Table position and scan planning are based on the correct patient reference.

Check PET / CT Workflow Alignment

Confirm whether the issue affects CT-only movement, PET-only bed indexing, or combined PET / CT acquisition.

PET / CT studies depend on accurate table position between CT attenuation correction and PET acquisition.

Expected outcome: Clinical Engineering can determine whether the positioning concern is isolated to one acquisition mode or affects combined system coordination.

Check for Recent Accessory or Table Configuration Changes

Ask whether any table pads, positioning accessories, patient supports, QA phantoms, injector setup, or immobilization devices were recently changed.

Make sure accessories are seated correctly and not shifting during table travel.

Expected outcome: A recent external setup change is identified or ruled out.

Power Cycle Only When Clinically Appropriate

If no patient is on the table and staff confirm it is safe, restart the operator workstation or system according to site practice.

Do not power cycle during an active patient scan unless directed by the clinical team and site policy.

Expected outcome: Temporary software or communication faults clear without unsafe interruption.

If table movement and indexing return to normal after restart, perform a safe functional check before returning the system to use.

Perform a Non-Patient Functional Movement Check

With no patient on the table, test basic table travel, height movement, load/unload position, and scan-range positioning if allowed by site procedure.

Observe for:

Expected outcome: The table moves smoothly and predictably without fault messages.

Check for Repeated or Intermittent Behavior

Ask whether the issue is constant or intermittent.

Intermittent table indexing or position feedback problems should not be dismissed, especially on PET / CT systems where accurate registration matters.

Expected outcome: The device status is based on repeatable observation, not a single unclear report.

Do Not Override or Force Table Movement

Do not push, pull, bypass, or mechanically force the patient table.

Do not continue scanning if table position accuracy is questionable.

Expected outcome: The system is not damaged further, and patient imaging accuracy is protected.

If the Problem Persists

If table movement, indexing, scan range positioning, or PET / CT alignment remains unreliable after external checks, common user-accessible causes have been ruled out.

The issue may involve internal table drive components, position feedback, motion control communication, calibration, encoder feedback, software coordination, or PET / CT registration-related service functions.

The device should be:

Knowing when to stop is proper troubleshooting. A PET / CT system with uncertain table position or scan range accuracy should not be used for patient imaging.

Clinical Use Tip

Do not troubleshoot table motion or scan range problems with an active patient on the table unless immediate safety requires it. Move the patient to a safe position first, then troubleshoot only when table movement will not create risk.

Do not troubleshoot on an active patient. Move the patient to a backup device or approved alternate imaging workflow first so therapy or diagnostic continuity is protected.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Technologist reported that the Siemens Biograph Horizon table stopped during PET bed indexing and the planned scan range did not advance correctly."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found no external obstruction, emergency stop activation, or accessory interference; table position fault repeated during non-patient movement check."

Resolution

What action was taken.

Example:
"Removed PET / CT system from service, labeled Out of Service, documented repeated table indexing fault, and escalated to Siemens service for evaluation."

Helpful Details to Include (If Known)

Final Thought

Patient table issues on a PET / CT system must be handled carefully because table position affects both patient safety and image registration accuracy. Start with external causes, verify workflow and positioning, avoid forcing movement, and escalate when position reliability cannot be confirmed. Clear CCR documentation helps service teams understand the failure and protects clinical operations.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide