Siemens Healthineers Biograph Vision

DICOM, PACS, or Worklist Transfer Fails

On this page

Asset Type

PET / CT System

Manufacturer

Siemens Healthineers

Model

Biograph Vision

What This Guide Helps With

Use this guide when worklist retrieval, DICOM transfer, or PACS communication fails because of network, destination, interface, or accessible configuration conditions.

Step-by-Step Troubleshooting

1. Preserve the Study and Maintain Clinical Workflow

Do not delete, recreate, or repeatedly manipulate patient studies solely to troubleshoot a communication problem.

Confirm that completed image data remains available locally and arrange an alternate approved workflow for urgent studies when necessary.

Expected outcome: Patient data is protected and clinical continuity is maintained.

2. Identify Which Communication Function Failed

Determine whether the issue affects:

Record any exact transfer or communication message.

Expected outcome: The failed communication path is clearly identified.

3. Confirm Local System Operation

Verify that the Biograph Vision workstation is otherwise operating normally and that studies can be accessed locally.

A local workstation or acquisition failure should be resolved before treating the issue as a network problem.

Expected outcome: Local scanner operation and study access are normal.

4. Inspect Accessible Network Connections

Check the scanner or workstation's accessible network cabling for:

Do not move the system to another port or network segment without authorization.

Expected outcome: Physical network connectivity appears intact.

If correcting an obvious external connection restores transfers, verify all required communication paths and stop.

5. Determine the Scope of the Network Problem

Check whether:

Coordinate with IT, PACS, or interface support when the problem is broader than the scanner.

Expected outcome: The failure is localized to the Biograph Vision, a specific destination, or shared infrastructure.

6. Verify Normal Operator-Accessible Destination Selection

Confirm that staff are using the intended configured destination and workflow.

Do not alter IP addresses, ports, AE Titles, routing, or protected network parameters unless specifically authorized and supported by approved documentation.

Expected outcome: The intended existing communication destination is selected.

If an incorrect normal destination selection caused the failure, correct the selection, verify transfer, and stop.

7. Retry One Representative Transfer

Once physical connectivity and destination availability are verified, retry one representative non-duplicative study transfer or approved test transaction.

Avoid repeatedly sending large batches before confirming that the communication path is functioning.

Expected outcome: The study transfers and is acknowledged by the intended destination.

If the transfer succeeds, verify receipt and stop.

8. Verify the Complete Communication Path

Confirm both sides of the workflow when possible:

A local β€œsent” indication alone should not be treated as proof of end-to-end success when confirmation is available.

Expected outcome: The intended end-to-end DICOM workflow functions correctly.

9. Perform Final Functional Verification

Verify all communication functions relevant to the reported problem, such as:

Expected outcome: The required DICOM communication path operates normally.

If all checks pass, troubleshooting is complete.

10. Escalate Persistent Communication Failure

If physical connectivity, destination availability, and normal workflow selection are correct but communication still fails, escalate to appropriate PACS, network, interface, or Siemens Healthineers support.

Expected outcome: The unresolved communication issue is routed to the team responsible for the remaining infrastructure or service-level fault.

If the Problem Persists

Common external causes have been ruled out. Remaining possibilities may involve network infrastructure, routing, DICOM services, PACS availability, interface engines, protected scanner network configuration, workstation software, or server-side conditions.

The device should be:

Do not make undocumented network changes in an attempt to restore connectivity.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

Confirm that urgent completed studies actually arrive at the receiving PACS; a scanner-side transfer attempt does not guarantee end-to-end delivery.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that completed Biograph Vision studies would not transfer to PACS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the workstation's accessible network cable loose at the wall connection."

Resolution

What action was taken.

Example:
"Clinical Engineering secured the network connection, successfully transferred a representative study, confirmed PACS receipt, and returned the communication workflow to service."

Helpful Details to Include (If Known)

Final Thought

Protect patient data, verify the physical network and end-to-end workflow before changing configuration, involve infrastructure teams when appropriate, and document exactly where communication failed and how it was verified afterward.

That is successful troubleshooting.